Accuracy Evaluation of the Omron HEM-7120 Automated Oscillometric Upper-arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060-2:2018 Numerical Criteria.
Tipanan Pisithkul, Kwanchai Pirojsakul
Division of Nephrology, Department of Pediatrics, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok, Thailand.
Key points
From the paper's conclusion- This study evaluated the Omron HEM-7120 automated oscillometric upper-arm home blood pressure monitor in children and adolescents against the numerical accuracy criteria of the AAMI/ESH/ISO Universal Standard.
Abstract
This study evaluated the Omron HEM-7120 automated oscillometric upper-arm home blood pressure monitor in children and adolescents against the numerical accuracy criteria of the AAMI/ESH/ISO Universal Standard. Test readings were compared with mercury auscultatory readings from two physicians using the same-arm sequential method. Three cuffs covered arm circumferences of 12-18, 17-22, and 22-32 cm. Of 90 participants, 85 were analyzed, yielding 255 comparisons. Mean differences were -4.0 ± 5.94 mmHg for systolic and -1.8 ± 7.20 mmHg for diastolic blood pressure, meeting Criterion 1. Participant standard deviations were 5.07 and 6.48 mmHg, below Criterion 2 limits. However, the requirements regarding age, reference-blood-pressure and cuff distributions, reference-measurement variability, and observer qualification were not fully met. The device met numerical criteria in this cohort, but it is not fully validated under the Universal Standard.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)Blood Pressure · 2026
- Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospective study.Blood pressure monitoring · 2026
- Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.Blood pressure monitoring · 2026
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Validation of the AViTA upper arm blood pressure monitor BPM636 in a combined pediatric and adult population according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2018/Amd2:2024.Blood pressure monitoring · 2026
- Validation of the novel professional automated auscultatory upper-arm cuff blood pressure monitor KOROT V2 Doctor (KOROT P2) in a pediatric population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018).Blood pressure monitoring · 2026