Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospective study.
HongYun Ma, JingZhen Liu, MengYa Chen, QianHui Yao
Nursing Department. · Cardiovascular Department, Luoyang Branch of Dongzhimen Hospital Affiliated to Beijing University of Chinese Medicine, Luoyang, Henan Province, China.
Key points
From the paper's conclusion- This study validated the accuracy and safety of the MED-58 upper-arm automated oscillometric blood pressure monitor - intended for home and clinical use - with fan-shaped and rigid cuffs, per International Organization for Standardization (ISO) 81060-2:2018 (Amd.1:2020; Amd.2:2024).
Abstract
This study validated the accuracy and safety of the MED-58 upper-arm automated oscillometric blood pressure monitor - intended for home and clinical use - with fan-shaped and rigid cuffs, per International Organization for Standardization (ISO) 81060-2:2018 (Amd.1:2020; Amd.2:2024). In this prospective, single-center validation study, participants aged ≥12 years were allocated to a fan-shaped cuff group (n = 85) or a rigid cuff group (n = 85). Per group, oscillometric readings with the MED-58 were compared with reference auscultatory measurements using the same-arm sequential method in ISO 81060-2. Blood pressure accuracy was assessed against ISO criterion 1 and 2. Secondary endpoints included pulse rate accuracy (±5% margin vs. manual palpation), device usability/performance, and safety. Both cuff configurations met ISO accuracy requirements. In the fan-shaped group, mean (±SD) device-reference differences were 1.1 ± 5.4 mmHg (SBP) and 0.3 ± 4.0 mmHg (DBP); SDs of per-participant mean differences were 4.4 mmHg (SBP) and 3.3 mmHg (DBP). In the rigid group, mean differences were 0.8 ± 5.2 mmHg (SBP) and 0.6 ± 4.5 mmHg (DBP); SDs of per-participant mean differences were 4.2 mmHg (SBP) and 3.8 mmHg (DBP). Mean pulse-rate differences were 1.65 bpm (fan-shaped) and 1.76 bpm (rigid), both within the prespecified ±5% equivalence margin. No adverse events were reported. The MED-58 demonstrated accuracy consistent with ISO 81060-2:2018 (Amd.1:2020; Amd.2:2024) with either cuff, with good usability and short-term safety.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)Blood Pressure · 2026
- Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.Blood pressure monitoring · 2026
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Validation of the LD-526 brachial blood pressure monitor according to the International Organization for Standardization protocol (ISO 81060-2:2018/AMD 1:2020/AMD 2:2024).Blood pressure monitoring · 2026
- Validation of the AViTA upper arm blood pressure monitor BPM636 in a combined pediatric and adult population according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2018/Amd2:2024.Blood pressure monitoring · 2026