Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)
Minghui Gong, Y S Yang, Yingying Chen, Jing Li, Jiayu Su, Yannan Huang, Kai Zhong, 张静 Jing ZHANG,
Dalian Medical University · Omron (Japan)
Key points
From the paper's conclusion- The Omron HEM-790XT1 automatic oscillometric upper-arm BP monitor satisfied all validation criteria of the AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020) standard and can be recommended for accurate BP measurement in adults within the validated arm circumference range.
Abstract
BACKGROUND: Validation of the accuracy of the Omron HEM-790XT1 automatic oscillometric upper-arm blood pressure (BP) monitor in the general adults according to the international AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020) standard. METHODS: Eligible adult participants meeting the protocol-specific criteria for age, gender, BP range and arm circumference distribution were recruited. The BP was measured by two trained observers using the same-arm sequential method. The test device measured BP during the inflation phase and was compared with readings from a standard mercury sphygmomanometer. The test device was used with its standard cuff, validated for an arm circumference range of 22-32 cm. RESULTS: A total of 85 participants were included in the final analysis after excluding 2 participants. The average age was 35 ± 13.7 years, with males accounting for 35.29% and females for 64.71%. According to Criterion 1 (all individual paired readings), the mean difference ± standard deviation (SD) between the test and reference device was -1.09 ± 4.94 mmHg for systolic blood pressure (SBP) and -2.49 ± 4.97 mmHg for diastolic blood pressure (DBP), meeting the requirement of ≤5 ± 8 mmHg. According to Criterion 2 (average per participant), the mean difference ± SD between the test and reference device was -1.09 ± 3.81 mmHg for SBP and -2.49 ± 4.12 mmHg for DBP. Both the SDs for SBP and DBP were below the derived thresholds of 6.86 and 6.47 mmHg, respectively. The Omron HEM-790XT1 automatic oscillometric BP monitor demonstrated satisfactory consistency with the mercury sphygmomanometer in both SBP and DBP measurements. CONCLUSIONS: The Omron HEM-790XT1 automatic oscillometric upper-arm BP monitor satisfied all validation criteria of the AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020) standard and can be recommended for accurate BP measurement in adults within the validated arm circumference range.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.Blood pressure monitoring · 2026
- Validation of the LD-526 brachial blood pressure monitor according to the International Organization for Standardization protocol (ISO 81060-2:2018/AMD 1:2020/AMD 2:2024).Blood pressure monitoring · 2026
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospective study.Blood pressure monitoring · 2026
- Accuracy Evaluation of the Omron HEM-7120 Automated Oscillometric Upper-arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060-2:2018 Numerical Criteria.Journal of clinical hypertension (Greenwich, Conn.) · 2026