Validation of the novel professional automated auscultatory upper-arm cuff blood pressure monitor KOROT V2 Doctor (KOROT P2) in a pediatric population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018).
Ariadni Menti, Aikaterini Theodosiadi, Vasileios Ntousopoulos, Eirini Chrysovalanti Antonogiannaki, Anastasios Kollias, George S Stergiou
Hypertension Center STRIDE-7, National and Kapodistrian University of Athens, School of Medicine, Third Department of Medicine, Sotiria Hospital, Athens, Greece.
Key points
From the paper's conclusion- The KOROT V2 Doctor (KOROT P2) device for professional use, which provides automated auscultatory measurements with visual display of the Korotkoff sounds, comfortably fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in a pediatric population and can be recommended for clinical use.
Abstract
OBJECTIVE: The novel automated auscultatory upper arm-cuff blood pressure (BP) monitor KOROT V2 Doctor (KOROT P2) has been successfully validated for professional use in the general population with three cuff sizes (23-28, 28-35, and 33-42 cm). The device provides automated measurements of BP using an electronic stethoscope embedded in the cuff, which records and displays Korotkoff sounds during deflation, allowing visual assessment by the healthcare professional. The device is also equipped with pediatric cuffs, and this study aimed to evaluate the device's accuracy in children. METHODS: The requirements of the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) were applied. Three cuffs of the test device were tested for arm circumference 15-18, 18-23, and 23-28 cm. RESULTS: Data from 35 individuals were analyzed [mean age: 8.7 ± 2.2 (SD) years, 18 boys, arm circumference: 15.1-28 cm]. For Validation Criterion 1, the mean difference ±SD between the test device and reference BP readings (N = 105) was 0.7 ± 5.1/0.4 ± 6.2 mmHg (systolic/diastolic; threshold ≤5 ± 8 mmHg). For Validation Criterion 2, the SD of the averaged BP differences between the test and reference BP readings per individual was 3.92/4.78 mmHg (systolic/diastolic; threshold ≤6.90/6.93 mmHg). CONCLUSION: The KOROT V2 Doctor (KOROT P2) device for professional use, which provides automated auscultatory measurements with visual display of the Korotkoff sounds, comfortably fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in a pediatric population and can be recommended for clinical use.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of the AViTA upper arm blood pressure monitor BPM636 in a combined pediatric and adult population according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2018/Amd2:2024.Blood pressure monitoring · 2026
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Accuracy Evaluation of the Omron HEM-7120 Automated Oscillometric Upper-arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060-2:2018 Numerical Criteria.Journal of clinical hypertension (Greenwich, Conn.) · 2026
- Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)Blood Pressure · 2026
- Clinical investigation of the iHealth Track KN-550BT upper-arm blood pressure monitor in pregnant patients according to the ISO 81060-2:2018+AMD1:2020/AMD2:2024.Blood pressure monitoring · 2026
- Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospective study.Blood pressure monitoring · 2026