Validation of the AViTA upper arm blood pressure monitor BPM636 in a combined pediatric and adult population according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2018/Amd2:2024.
Chung-Yu Lin, Yu-Hsuan Chien, I-Chih Huang
Department of Cardiology. · Division of Pediatrics, Fu Jen Catholic University Hospital.
Key points
From the paper's conclusion- The AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) protocol. It can be recommended for home-based blood pressure measurements in the tested population, including children aged 5-12 years and adults aged 21-82 years, with arm circumferences ranging from 15 to 42 cm.
Abstract
OBJECTIVE: To evaluate the clinical accuracy of the AViTA BPM636 oscillometric upper-arm blood pressure monitor in the tested pediatric and adult population, including children aged 5-12 years and adults aged 21-82 years with arm circumferences ranging from 15 to 42 cm, according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) Universal Standard (ISO 81060-2:2018/Amd2:2024). METHODS: The AViTA BPM636 was evaluated in 85 participants (including 35 pediatric and 50 adult subjects) with upper-arm circumferences ranging from 15 to 42 cm, and two cuffs of different sizes (15 to 22 and 22 to 42 cm) were used in this study. The recruitment fulfilled the requirements of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) for a combined pediatric and adult population using the same arm sequential blood pressure measurement method. RESULTS: For criterion 1, the mean ± SD of the differences between the test device and reference blood pressure measurements was 0.93 ± 6.11 mmHg for SBP and -0.20 ± 5.64 mmHg for DBP. For criterion 2, the SD of individual mean differences for SBP (5.44 mmHg) and DBP (4.98 mmHg) remained well within the maximum permissible limits of 6.88 and 6.95 mmHg, respectively, thereby satisfying the validation criteria. CONCLUSION: The AViTA BPM636 passed the criteria of the ANSI/AAMI/ISO Universal Standard (ISO 81060-2:2018/Amd2:2024) protocol. It can be recommended for home-based blood pressure measurements in the tested population, including children aged 5-12 years and adults aged 21-82 years, with arm circumferences ranging from 15 to 42 cm.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Validation of the LD-526 brachial blood pressure monitor according to the International Organization for Standardization protocol (ISO 81060-2:2018/AMD 1:2020/AMD 2:2024).Blood pressure monitoring · 2026
- Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)Blood Pressure · 2026
- Validation of the novel professional automated auscultatory upper-arm cuff blood pressure monitor KOROT V2 Doctor (KOROT P2) in a pediatric population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018).Blood pressure monitoring · 2026
- Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospective study.Blood pressure monitoring · 2026
- Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.Blood pressure monitoring · 2026