Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.

Xiaojing Guo

一句话结论 · In one sentence

The Arm-Type Fully Automatic Blood Pressure Monitor (model: DBP-62F4B-P) meets the accuracy and consistency requirements of the AAMI/ESH/ISO (ISO 81060-2:2018+Amd 2:2024) standard for SBP and DBP measurement in adolescents and adults with arm circumference 22.0-42.0 cm. It can be safely and effectively used for BP monitoring in clinical settings and, by extension, is suitable for home use based on its design features and validation performance.

原始摘要(英文原文)· Original abstract
OBJECTIVE: To validate the accuracy and safety of the Arm-Type Fully Automatic Blood Pressure Monitor (model: DBP-62F4B-P) manufactured by JOYTECH Healthcare Co., Ltd. for blood pressure (BP) measurement in adolescents and adults with arm circumference 22.0-42.0 cm, in accordance with the Association for the Advancement of Medical Instrumentation (AAMI)/European Society of Hypertension (ESH)/International Organization for Standardization (ISO) (ISO 81060-2:2018+Amd 2:2024) international standard. METHODS: A total of 90 subjects were enrolled. Three valid datasets were collected from each subject, yielding 270 datasets for analysis. The mercury sphygmomanometer (reference device) data were used as the gold standard to evaluate whether the accuracy of the test device met the requirements. RESULTS: Based on 270 valid datasets, the mean difference between the test device and the reference device for SBP was -0.49 mmHg (SD = 7.29 mmHg), and for DBP was 1.45 mmHg (SD = 6.14 mmHg), meeting Criterion 1 requirements. Analysis based on individual averages of the 90 subjects showed a mean difference of -0.49 mmHg (SD = 5.57 mmHg) for SBP and 1.45 mmHg (SD = 5.05 mmHg) for DBP, both meeting the requirements of Criterion 2. The device maintained stable measurement performance across different arm circumferences, ages, and BP levels. CONCLUSION: The Arm-Type Fully Automatic Blood Pressure Monitor (model: DBP-62F4B-P) meets the accuracy and consistency requirements of the AAMI/ESH/ISO (ISO 81060-2:2018+Amd 2:2024) standard for SBP and DBP measurement in adolescents and adults with arm circumference 22.0-42.0 cm. It can be safely and effectively used for BP monitoring in clinical settings and, by extension, is suitable for home use based on its design features and validation performance.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard. — 科研速览 Science Skim