Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.
Yonghua Qiu, Wei Han, Lijun Liu, Huirong Chen, Zhiping Sun, Kexin Zhu
Department of Cardiology, Foshan Fosun Chancheng Hospital, Foshan, Guangdong Province, China.
Key points
From the paper's conclusion- This study evaluated the overall effectiveness and safety of the ARM-30F-XL arm blood pressure (BP) monitor in individuals with an arm circumference of 22-52 cm using a mercury sphygmomanometer as the reference device.
- The test device was the ARM-30F-XL, and the reference device was a mercury sphygmomanometer. Using BP readings obtained from the mercury sphygmomanometer as the reference standard, the accuracy of the noninvasive BP module of the test device was evaluated to determine whether it met the specified requirements.
- The ARM-30F-XL complies completely with International Standard ISO 81060-2:2018+Amd 2:2024 and is recommended for BP measurement in individuals with an arm circumference of 22-52 cm for home and clinical use.
Abstract
To evaluate the overall effectiveness and safety of the ARM-30F-XL arm blood pressure (BP) monitor in individuals with an arm circumference of 22-52 cm. Subjects were recruited from patients, patient companions, and individuals attending routine health examinations at the hospital. The test device was the ARM-30F-XL, and the reference device was a mercury sphygmomanometer. Using BP readings obtained from the mercury sphygmomanometer as the reference standard, the accuracy of the noninvasive BP module of the test device was evaluated to determine whether it met the specified requirements. This study analyzed valid BP data from 97 subjects. According to criterion 1, the mean difference in SBP between the test device and the reference device was 1.69 mmHg, and the SD was 5.00 mmHg. The mean difference in DBP was -2.16 mmHg, and the SD was 4.45 mmHg. The mean difference of both SBP and DBP was less than ±5 mmHg, and the SD was less than 8 mmHg, which met the requirements. According to criterion 2, the SD of SBP was 4.00 mmHg, which was less than 6.73 mmHg and met the requirements. The SD of DBP was 3.73 mmHg, which was less than 6.58 mmHg and met the requirements. The ARM-30F-XL complies completely with International Standard ISO 81060-2:2018+Amd 2:2024 and is recommended for BP measurement in individuals with an arm circumference of 22-52 cm for home and clinical use.
Read the paper
Discussion · 1
Science Skim's AI key points, plus readers' public thoughts
- This study evaluated the overall effectiveness and safety of the ARM-30F-XL arm blood pressure (BP) monitor in individuals with an arm circumference of 22-52 cm using a mercury sphygmomanometer as the reference device.
- The test device was the ARM-30F-XL, and the reference device was a mercury sphygmomanometer. Using BP readings obtained from the mercury sphygmomanometer as the reference standard, the accuracy of the noninvasive BP module of the test device was evaluated to determine whether it met the specified requirements.
- The ARM-30F-XL complies completely with International Standard ISO 81060-2:2018+Amd 2:2024 and is recommended for BP measurement in individuals with an arm circumference of 22-52 cm for home and clinical use.
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)Blood Pressure · 2026
- Validation of the LD-526 brachial blood pressure monitor according to the International Organization for Standardization protocol (ISO 81060-2:2018/AMD 1:2020/AMD 2:2024).Blood pressure monitoring · 2026
- Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospective study.Blood pressure monitoring · 2026
- Clinical validation of MESI mTABLET BP according to ISO 81060-2:2018 + A1 + A2.Blood pressure monitoring · 2026
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026