Validation of the LD-526 brachial blood pressure monitor according to the International Organization for Standardization protocol (ISO 81060-2:2018/AMD 1:2020/AMD 2:2024).
Yi Zhou, Xin-Yu Wang, Ruo-Yi Zou, Ji-Guang Wang
Department of Cardiovascular Medicine, State Key Laboratory of Medical Genomics, Shanghai Key Laboratory of Hypertension, Shanghai Institute of Hypertension, National Research Centre for Translational Medicine, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Key points
From the paper's conclusion- The LD-526 BP monitor fulfilled all the requirements of the ISO universal standard (ISO 81060-2 : 2018/AMD 1 : 2020/AMD 2 : 2024) in the general population and can be recommended for self-measurement.
Abstract
OBJECTIVE: The LD-526 oscillometric upper-arm blood pressure (BP) monitor is intended for self-measurement of BP in adults. The aim of this study was to evaluate the accuracy of LD-526 oscillometric upper-arm BP monitor in the general population according to the International Organization for Standardization (ISO) universal standard (ISO 81060-2 : 2018/AMD 1 : 2020/AMD 2 : 2024). METHOD: Subjects were recruited to fulfill the age, gender, BP, and cuff distributions of the ISO universal standard in the general population using the same-arm sequential BP measurement method. The test device was used with one universal cuff for arm circumferences of 22-42 cm. Reference BP measurements were obtained using a calibrated mercury sphygmomanometer, with two cuffs (inflatable bladder sizes of 12 × 24 and 16 × 32 cm, respectively). RESULTS: Of the 98 recruited subjects, 85 [women/men: 46/39; mean (±SD) age: 52.2 ± 17.4 years] were analyzed, and 13 were excluded from the final analysis: 10 because of reference BP variability and three because of cardiac arrhythmias. For validation of criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 1.34 ± 5.56/-0.31 ± 4.36 mmHg (systolic/diastolic). For criterion 2, the SD of the average BP differences between the test device and reference BP per subject was 4.51/3.77 mmHg (systolic/diastolic). CONCLUSION: The LD-526 BP monitor fulfilled all the requirements of the ISO universal standard (ISO 81060-2 : 2018/AMD 1 : 2020/AMD 2 : 2024) in the general population and can be recommended for self-measurement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)Blood Pressure · 2026
- Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.Blood pressure monitoring · 2026
- Validation of the AViTA upper arm blood pressure monitor BPM636 in a combined pediatric and adult population according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization 81060-2:2018/Amd2:2024.Blood pressure monitoring · 2026
- Clinical validation of MESI mTABLET BP according to ISO 81060-2:2018 + A1 + A2.Blood pressure monitoring · 2026
- Clinical validation of the MED-58 upper-arm automated oscillometric blood pressure monitor with fan-shaped and rigid cuffs for home and clinical use in adolescents and adults, according to International Organization for Standardization 81060-2:2018 (Amd.1:2020; Amd.2:2024): a single-center prospective study.Blood pressure monitoring · 2026