Validation of the LD-526 brachial blood pressure monitor according to the International Organization for Standardization protocol (ISO 81060-2:2018/AMD 1:2020/AMD 2:2024).

Yi Zhou, Xin-Yu Wang, Ruo-Yi Zou, Ji-Guang Wang

Department of Cardiovascular Medicine, State Key Laboratory of Medical Genomics, Shanghai Key Laboratory of Hypertension, Shanghai Institute of Hypertension, National Research Centre for Translational Medicine, Ruijin Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China.

Key points

From the paper's conclusion
  • The LD-526 BP monitor fulfilled all the requirements of the ISO universal standard (ISO 81060-2 : 2018/AMD 1 : 2020/AMD 2 : 2024) in the general population and can be recommended for self-measurement.

Abstract

OBJECTIVE: The LD-526 oscillometric upper-arm blood pressure (BP) monitor is intended for self-measurement of BP in adults. The aim of this study was to evaluate the accuracy of LD-526 oscillometric upper-arm BP monitor in the general population according to the International Organization for Standardization (ISO) universal standard (ISO 81060-2 : 2018/AMD 1 : 2020/AMD 2 : 2024). METHOD: Subjects were recruited to fulfill the age, gender, BP, and cuff distributions of the ISO universal standard in the general population using the same-arm sequential BP measurement method. The test device was used with one universal cuff for arm circumferences of 22-42 cm. Reference BP measurements were obtained using a calibrated mercury sphygmomanometer, with two cuffs (inflatable bladder sizes of 12 × 24 and 16 × 32 cm, respectively). RESULTS: Of the 98 recruited subjects, 85 [women/men: 46/39; mean (±SD) age: 52.2 ± 17.4 years] were analyzed, and 13 were excluded from the final analysis: 10 because of reference BP variability and three because of cardiac arrhythmias. For validation of criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 1.34 ± 5.56/-0.31 ± 4.36 mmHg (systolic/diastolic). For criterion 2, the SD of the average BP differences between the test device and reference BP per subject was 4.51/3.77 mmHg (systolic/diastolic). CONCLUSION: The LD-526 BP monitor fulfilled all the requirements of the ISO universal standard (ISO 81060-2 : 2018/AMD 1 : 2020/AMD 2 : 2024) in the general population and can be recommended for self-measurement.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers