Clinical investigation of the iHealth Track KN-550BT upper-arm blood pressure monitor in pregnant patients according to the ISO 81060-2:2018+AMD1:2020/AMD2:2024.
Ying Zhang, Dongmei Wang, Chunmei Cui, Yu Zhang, Jieqiong Zhang, Yi Wan
Department of Gynaecology and Obstetrics, Xijing 986 Hospital, Fourth Military Medical University, Xi'an. · Department of Gynaecology and Obstetrics, Central Theater Command Air Force Hospital, Datong.
Key points
From the paper's conclusion- The iHealth Track KN-550BT upper-arm oscillometric automated BP monitor fully complies with the requirements of ISO 81060-2:2018+AMD1:2020/AMD2:2024, and can be recommended for clinical application and home self-BP measurement among pregnant patients.
Abstract
OBJECTIVE: Validation of blood pressure (BP) devices is essential for reliable BP detection in pregnant populations. This study, therefore, set out to assess the measurement performance of the iHealth Track KN-550BT oscillometric upper-arm automated BP monitor for pregnant patients with the standard International Organization for Standardization (ISO) 81060-2:2018+AMD1:2020/AMD2:2024. METHODS: This validation study enrolled a total of 45 pregnant patients. All BP measurements were performed sequentially on the left arm using the ISO unified standardized protocol. Device-reference discrepancies between the tested automated sphygmomanometer and the mercury sphygmomanometer were calculated and evaluated in strict accordance with the ISO standard specifications. RESULTS: The study cohort consisted of 15 normotensive pregnant patients, 15 patients with gestational hypertension, and 15 patients diagnosed with preeclampsia. For the ISO Validation Criterion 1, the mean differences between the test device and mercury standard were -0.90 ± 4.82 mmHg for SBP and -0.91 ± 4.14 mmHg for DBP, respectively. For Criterion 2, the standard deviations of the averaged per-participant differences between device and reference measurements were 3.95 mmHg for SBP and 3.10 mmHg for DBP, respectively. CONCLUSION: The iHealth Track KN-550BT upper-arm oscillometric automated BP monitor fully complies with the requirements of ISO 81060-2:2018+AMD1:2020/AMD2:2024, and can be recommended for clinical application and home self-BP measurement among pregnant patients.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Update of the practical guide for validation studies of automated cuff blood pressure monitors according to the Universal Standard by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018) and Amendments 1:2020 and 2:2024.Journal of hypertension · 2026
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026
- Validation of the LD-526 brachial blood pressure monitor according to the International Organization for Standardization protocol (ISO 81060-2:2018/AMD 1:2020/AMD 2:2024).Blood pressure monitoring · 2026
- Validating the accuracy of Omron HEM-790XT1 electronic blood pressure monitor in the general population according to AAMI/ESH/ISO (ISO 81060-2:2018 + AMD1:2020)Blood Pressure · 2026
- Validation of the Arm-Type Fully Automatic Blood Pressure Monitor DBP-62F4B-P in adolescents and adults according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (ISO 81060-2:2018+Amd 2:2024) standard.Blood pressure monitoring · 2026
- Validation of the ARM-30F-XL arm blood pressure monitor with multiple cuff sizes according to the International Standard ISO 81060-2:2018+Amd 2:2024.Blood pressure monitoring · 2026