Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.
Farhad Ahamed, Debkumar Pal, Sarika Palepu, Jeevanmuktha Somashekara, Sibasish Sahoo, Ayan Roy, Tanmay Saha, Semanti Das,
Department of Community Medicine and Family Medicine, All India Institute of Medical Sciences, Kalyani, West Bengal, India. · Department of Biochemistry, All India Institute of Medical Sciences, Kalyani, West Bengal, India.
Key points
From the paper's conclusion- The POCT device showed substantial concordance with the gold standard method for measuring both FBS and PPBS. It also demonstrated a fair agreement with the automated analyzer in estimating blood glucose levels.
Abstract
INTRODUCTION: We aimed to conduct this study to compare the validity of the HemoCue® Glucose 201 RT System (point-of-care testing (POCT) device) in estimating blood glucose levels in patients attending a tertiary care hospital, considering the automated analyzer method (Vitros® 4600 chemistry system, QuidelOrtho Corporation, USA) as the gold standard. METHODS: We conducted this study among the patients attending the outpatient department in a tertiary care hospital in Eastern India. Venous blood was collected to estimate fasting blood glucose (FBS) and postprandial blood glucose (PPBS) using the automated analyzer and the POCT device, along with capillary blood to estimate FBS and PPBS using the POCT device. RESULTS: We enrolled 200 patients with a mean age of 49.4 (standard deviation 12.6) years. Lin's CCC for POCT device compared to automated analyzer (venous blood) for FBS estimation was 0.95 (95% confidence interval (CI): 0.94 to 0.96, P value < 0.001) for venous blood and 0.95 (95% CI: 0.94-0.97, P value < 0.001) for capillary blood. Lin's CCC for PPBS estimation was 0.99 (95% CI: 0.98-0.99, P value < 0.001) for venous blood and 0.98 (95% CI: 0.97-0.98, P value < 0.001) for capillary blood. CONCLUSION: The POCT device showed substantial concordance with the gold standard method for measuring both FBS and PPBS. It also demonstrated a fair agreement with the automated analyzer in estimating blood glucose levels.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative Performance Analysis of Three Point-of-Care Glucose Testing Devices against a Central Laboratory Reference Method.EJIFCC · 2026
- Point-of-care glucose testing (POCT) replacing venous blood sampling in gestational diabetes mellitus screening; a comparative study.BMC pregnancy and childbirth · 2026
- A-378 Performance Evaluation of the Roche Cobas® Pulse Point-of-Care GlucometerClinical Chemistry · 2026
- Point-of-Care Quantification of HbA1c and CRP using Finecare: Analytical Validation of a Fluorescence-Based System.Annals of clinical biochemistry · 2026
- The utility of point-of-care haemoglobin A1c tests in a very remote Aboriginal and Torres Strait Islander community: an agreement analysis with routine laboratory testing.Rural and remote health · 2026
- A-386 Evaluation of near patient beta hydroxybutyrate testing in post mortem samples using the StatStrip analyserClinical Chemistry · 2026