The utility of point-of-care haemoglobin A1c tests in a very remote Aboriginal and Torres Strait Islander community: an agreement analysis with routine laboratory testing.
Hasthi Udyangani Wanamali Dissanayake, Sabine Braat, George Gurruwiwi, Ruth Gulamanda, Josh C Tynan, Lindy Marlow, Benjamin Harrap, Kenny Gongnini,
Department of Infectious Diseases, Peter Doherty Institute for Infection and Immunity, The University of Melbourne, Melbourne, Vic. 3000, Australia. · Yardhura Walani, National Centre for Epidemiology and Population Health, Australian National University, Canberra, ACT 2601, Australia.
Key points
From the paper's conclusion- Capillary measurement of HbA1c using a portable point-of-care device demonstrated promising agreement with routine laboratory-tested venous blood HbA1c. This supports the development of novel outreach models of diabetes care that meet remote Aboriginal and Torres Strait Islander community needs.
Abstract
INTRODUCTION: Optimal blood glucose control is critical to the prevention of diabetes complications. Point-of-care testing (POCT) provides a flexible alternative to conventional approaches that rely on laboratory processing of venous blood samples, with variable time to receiving results. We determined the level of agreement between the A1CNow+ glycated haemoglobin POCT system and routine laboratory haemoglobin A1C (HbA1c) testing in a remote Aboriginal Australian community. METHODS: HbA1c was measured in Aboriginal and Torres Strait Islander adults using capillary blood and the point-of-care device and venous blood using standard laboratory procedures. Agreement between the point-of-care and laboratory results was examined using a modified Bland-Altman analysis and diagnostic agreement for type 2 diabetes using Cohen's kappa coefficient. Capillary and venous blood collection were conducted in a clinical setting. RESULTS: Participants (n=49) were on average age of 43±10 years, 65% were female, and with mean BMI of 30±5 kg/m2. Median HbA1c was 5.7% (39 mmol/mol) for both methods of measurement. The mean difference between the A1CNow+ point-of-care test and laboratory HbA1c measurement was 0.03% (0.34 mmol/mol), with limits of agreement of -0.75 to 0.81% (-8.15-8.83 mmol/mol). Diagnostic agreement for type 2 diabetes (HbA1c ≥6.5% (48 mmol/mol)) was moderate (Cohen's kappa statistic 0.70), with minor discrepancies observed at higher HbA1c levels. DISCUSSION: These findings support the clinical usefulness of portable HbA1c point-of-care testing beyond traditional clinic settings. While clinic-based models have improved diabetes care, they do not reach all remote communities effectively. Home-based outreach health care utilising A1CNow+ point-of-care testing may improve access to care, strengthen family engagement, and reduce cultural and logistical barriers. Despite sparse data at HbA1c levels greater than 55 mmol/mol, overall benefits appear meaningful. CONCLUSION: Capillary measurement of HbA1c using a portable point-of-care device demonstrated promising agreement with routine laboratory-tested venous blood HbA1c. This supports the development of novel outreach models of diabetes care that meet remote Aboriginal and Torres Strait Islander community needs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.Indian journal of community medicine : official publication of Indian Association of Preventive & Social Medicine · 2026
- Scoping review of HbA1c device use in selected Asian countries: Implications for point-of-care testing in Nepal.PloS one · 2026
- Point-of-care glucose testing (POCT) replacing venous blood sampling in gestational diabetes mellitus screening; a comparative study.BMC pregnancy and childbirth · 2026
- Mobile point-of-care testing to improve access to care for people experiencing homelessness and diabetes in Los Angeles.American journal of clinical pathology · 2026
- Point-of-Care Quantification of HbA1c and CRP using Finecare: Analytical Validation of a Fluorescence-Based System.Annals of clinical biochemistry · 2026
- Method comparison and agreement of a fluorescence immunoassay-based HbA1c test with a routine HPLC method.Diabetology international · 2026