A-386 Evaluation of near patient beta hydroxybutyrate testing in post mortem samples using the StatStrip analyser
Stephen Morley, Lucy Ratajczak, Lynsey Burke
Key points
From the paper's conclusion- Conclusion The StatStrip analyser is not suitable for NPT OHB measurement.
Abstract
Background Diabetic and alcoholic ketoacidosis is a common cause of death. If a pathologist could use a near-patient testing (NPT) method to screen for betahydroxybutyrate (OHB), and so ketoacidosis this would allow for a more focused approach to an autopsy. There are already validated NPT methods for OHB used in the living. This study compared the StatStrip analyser (Nova Biomedical) against the in validated house laboratory method (spectrophotometric). Noba Biomedical do not marker the analyser for post mortem work. In post mortem work it is well established that the blood and vitreous OHB are comparable Methods 25 cases in which OHB had been measured in blood were investigated. There were 25 cases in which there was sufficient blood provided and 9 cases in which there was a concurrent vitreous sample. The blood was initially tested using the validated in-house spectrophotometric method. The blood and vitreous were then tested using the StatStrip analyser and result compared. Results 11 of the 25 (44%) of blood and 2 of the 9 (22%) of the vitreous had flow errors. In blood, the 14 cases with comparable results had a poor correlation (R2 = 0.54), and all cases had a significant negative bias with the worst StatStrip result of 0.1 mmol/l compared to an in house reading of 7.99 mmol/L. The comparison of the validated blood method to vitreous was also poor, (R2 = 0.0491) with a maximum discrepancy of 7.99 mmol/L in the validated blood method and 0.5 mmol/l in vitreous. Conclusion The StatStrip analyser is not suitable for NPT OHB measurement. This is likely due to the flow dynamics required in lateral flow testing.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparing clinical laboratory beta-hydroxybutyrate testing of vitreous humor against toxicological blood analyses in ketoacidosis deaths.Journal of forensic sciences · 2026
- A-360 Performance of rapid glucose and ketone screening methods in post-mortem vitreous fluid samples to determine ketoacidosis-mediated deathClinical Chemistry · 2026
- A-159 Specificity of beta hydroxybutyrate in the confirmation of insulin-induced hypoglycemia in post mortem casesClinical Chemistry · 2026
- B-406 The reliability of a significantly raised acetone (>10mg/100mL blood) to diagnose ketoacidosis in post mortem caseworkClinical Chemistry · 2026
- B-028 Review of a sample exchange program for postmortem vitreous biochemical analysisClinical Chemistry · 2026
- Analytical Validation of β-Hydroxybutyrate Quantification in Feline Serum with the BT3500 Analyzer.Veterinary clinical pathology · 2026