Comparative Performance Analysis of Three Point-of-Care Glucose Testing Devices against a Central Laboratory Reference Method.
Auliya Ashar, Fauqa Arinil Aulia, Ferdy Royland Marpaung, M Robiul Fuadi, Yessy Puspitasari
Clinical Pathology Study Program, Faculty of Medicine, Universitas Airlangga, Surabaya, Indonesia. · Department of Clinical Pathology, Faculty of Medicine, Universitas Airlangga, Surabaya, Indonesia.
Key points
From the paper's conclusion- Although all POCT systems showed strong correlations with the reference method, clinically relevant differences in analytical agreement and accuracy were identified. Rightest demonstrated the best overall performance. These findings highlight the importance of evaluating agreement beyond correlation and support the use of POCT primarily for glucose monitoring rather than definitive diagnosis.
Abstract
BACKGROUND: Point-of-care testing (POCT) glucose systems are widely used in clinical practice due to their rapid turnaround time and operational convenience. However, concerns remain regarding their analytical agreement with laboratory reference methods, particularly in relation to accuracy and potential clinical impact. METHODS: This cross-sectional analytical study included 196 adult outpatients at Dr. Soetomo General Academic Hospital, Surabaya, Indonesia. Capillary blood glucose was measured using three POCT devices (StatStrip, Accu-Chek Inform II, and Rightest), while venous plasma glucose was analyzed using a hexokinase-based reference method (Alinity C). Analytical performance was evaluated using correlation analysis, Bland-Altman plots, Passing-Bablok regression, and ISO 15197 accuracy criteria. RESULTS: All POCT devices demonstrated strong correlations with the reference method (r = 0.967-0.982; p < 0.001). StatStrip showed a small negative bias (-2.89 mg/dL; p = 0.003), while Accu-Chek Inform II exhibited a larger deviation (-7.80 mg/dL; p < 0.001). In contrast, Rightest showed no statistically significant bias (-0.14 mg/dL; p = 0.870) and demonstrated the closest agreement with the reference method. Agreement analysis confirmed minimal systematic and proportional bias for Rightest, whereas Accu-Chek Inform II showed wider variability. ISO 15197 evaluation indicated the highest accuracy for Rightest and lower compliance for Accu-Chek Inform II, particularly at glucose levels <200 mg/dL. CONCLUSION: Although all POCT systems showed strong correlations with the reference method, clinically relevant differences in analytical agreement and accuracy were identified. Rightest demonstrated the best overall performance. These findings highlight the importance of evaluating agreement beyond correlation and support the use of POCT primarily for glucose monitoring rather than definitive diagnosis.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.Indian journal of community medicine : official publication of Indian Association of Preventive & Social Medicine · 2026
- Comparison of Blood Glucose Measurement Methods at Low Concentrations in Neonates at Risk for Hypoglycemia.The journal of applied laboratory medicine · 2026
- A-378 Performance Evaluation of the Roche Cobas® Pulse Point-of-Care GlucometerClinical Chemistry · 2026
- Analytical Agreement and Diagnostic Performance of a Commercial Point-of-Care Lipid Testing System for Lipid Profile and Atherogenic Ratios.Diagnostics (Basel, Switzerland) · 2026
- Point-of-Care Quantification of HbA1c and CRP using Finecare: Analytical Validation of a Fluorescence-Based System.Annals of clinical biochemistry · 2026
- A-404 Determinants of Analytical Performance in Centrifugation–Based Hematocrit: Effects of Calibration and Internal Quality Control Frequency Across Hospital and Primary Point-of-Care SitesClinical Chemistry · 2026