Point-of-Care Quantification of HbA1c and CRP using Finecare: Analytical Validation of a Fluorescence-Based System.
Ali Khatib, Álvaro Martínez Ortega, Noor Lachhab, Iman Azariouh Kaddouri, Soukayna Draoui, Jacques Vanderlinden, Damien Gruson
Université catholique de Louvain. · Service of Clinical AnalysisHospital Universitario La Paz.
Key points
From the paper's conclusion- The Finecare™ POCT device demonstrated acceptable analytical performance for CRP in acute care. However, due to elevated imprecision and workflow complexity for HbA1c, its use should be limited to settings prioritizing rapid decision-making over maximum precision.
Abstract
BACKGROUND: This study evaluated the analytical performance and practicability of a novel point-of-care testing (POCT) device for measuring HbA1c and C-reactive protein (CRP) compared with central laboratory methods. METHODS: Imprecision, bias, and analytical total error (TEa) were evaluated using quality control materials over five days (n=20 per level). Method comparison was performed on 120 patient venous samples per biomarker. Practicality was assessed using an operations scorecard. RESULTS: CRP performance met EFLM biological variation specifications, with TEa values (25.95% and 18.57%) remaining below desirable limits (TEa < 38.14%). Method comparison showed a strong correlation (r = 0.98), despite proportional bias at low levels (< 20.0 mg/L), and high categorical agreement (κ = 0.92). For HbA1c, TEa exceeded EFLM minimum limits (TEa < 3.56%) and central laboratory goals (TEa < 6.0%). At Control Level 1, TEa (10.28%) aligned with CLIA POCT standards (TEa < 10%), but reached 14.75% at Control Level 2 due to elevated imprecision. Slight proportional bias was observed at low HbA1c levels (< 7.0%), though categorical agreement remained robust (κ = 0.90). Practicability evaluation indicated general user satisfaction, although HbA1c testing was perceived as more complex. CONCLUSION: The Finecare™ POCT device demonstrated acceptable analytical performance for CRP in acute care. However, due to elevated imprecision and workflow complexity for HbA1c, its use should be limited to settings prioritizing rapid decision-making over maximum precision.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.Indian journal of community medicine : official publication of Indian Association of Preventive & Social Medicine · 2026
- Evaluation of the Lifotronic H100 Plus analyzer for HbA1c determination.Scandinavian journal of clinical and laboratory investigation · 2026
- Method comparison and agreement of a fluorescence immunoassay-based HbA1c test with a routine HPLC method.Diabetology international · 2026
- Evaluation of point-of-care complete blood count and C-reactive protein testing for routine use in a pediatric emergency department.European journal of pediatrics · 2026
- Comparative Performance Analysis of Three Point-of-Care Glucose Testing Devices against a Central Laboratory Reference Method.EJIFCC · 2026
- Evaluation of Point-of-Care Testing for Creatinine, Estimated Glomerular Filtration Rate, and Potassium in a Diverse United Kingdom Population.The journal of applied laboratory medicine · 2026