A-378 Performance Evaluation of the Roche Cobas® Pulse Point-of-Care Glucometer
Pit Hung Lim, Sian Foong Lim, Kelly Lim, Lizhen Ong
Key points
From the paper's conclusion- Conclusion The Roche cobas® pulse glucometer demonstrated excellent analytical performance, meeting all manufacturer's claims, RCPA performance specifications, FDA hospital use recommendation and CAP evaluation criteria.
Abstract
Background Point-of-care testing (POCT) for glucose monitoring plays a crucial role in enabling timely clinical decisions within healthcare settings. The Roche cobas® pulse is a POCT device for glucose measurement and is compatible with cobas® GLU test strips and cobas® GLU QC kit for internal quality control (Roche Diagnostics GmbH). This new generation glucometer utilises FAD-dependent glucose dehydrogenase (GDH) enzyme technology and is intended to replace the current Roche Inform II system. This study provided a real-world analytical and clinical accuracy evaluation of the Roche cobas® pulse glucometer compared to its predecessor Roche Inform II. Methods Between June to October 2025, we evaluated Roche cobas® pulse glucometer (FAD-GDH electrochemical method) against its predecessor Roche Inform II (mutant quinoprotein GDH electrochemical method) and referencing to central laboratory platform testing on Abbott CI hexokinase glucose method for method comparison, imprecision, linearity and accuracy. Within-day and between-day imprecision were assessed using two QC levels (Low level: 3.3 mmol/L; High level: 17.7 mmol/L). Linearity was verified using commercial linearity material ranging from 1.5 to 29.7 mmol/L. Method correlation studies were performed using 77 patient samples for each comparison (Alinity CI and Inform II) across the reportable range of 0.6 to 33.3 mmol/L. Accuracy was verified using CAP WBGQ-B 2025 EQA samples. Performance was evaluated using Royal College of Pathologists of Australasia (RCPA) allowable performance specifications (±0.4 mmol/L for glucose =5.0 mmol/L or ±8% for glucose >5.0 mmol/L) and Food and Drug Administration (FDA) recommendations of 98% accuracy within ±15% for glucose = 4.2 mmol/L or ±0.83 mmol/L for glucose < 4.2 mmol/L. Results The Roche cobas® pulse showed excellent precision with within-day CV of 1.0% and 0.5% and between-day CV of 1.3% and 0.4% for low and high levels respectively, fulfilling manufacturer claims. Linearity was verified across the tested range (R²=0.9998). Correlation with Abbott Alinity CI showed strong agreement (cobas® pulse = 0.9291[Alinity CI] + 0.2634, R²=0.9979, n=76), with 71 of 76 samples (93.4%) meeting RCPA allowable performance specifications (APS) after excluding 1 outlier. Comparison with Roche Inform II showed excellent correlation (cobas® pulse = 0.9693[Inform II] + 0.013, R²=0.9940, n=77), with 73 of 77 samples (94.8%) meeting RCPA APS. Both comparisons achieved 100% compliance with FDA hospital-use accuracy criteria, exceeding the 98% requirement. Accuracy verification using CAP EQA samples showed all results within ±2 SDI and CAP evaluation criteria (peer group mean ±12.5%). The Roche cobas® pulse glucometer demonstrated analytical equivalence to Inform II, with improved precision, equivalent linearity across the clinical range and comparable accuracy. Conclusion The Roche cobas® pulse glucometer demonstrated excellent analytical performance, meeting all manufacturer's claims, RCPA performance specifications, FDA hospital use recommendation and CAP evaluation criteria. It performed comparably to Roche Inform II and to central laboratory platform on Abbott Alinity, >93% of samples within RCPA APS criteria, confirms its suitability for replacing the Roche Inform II glucometer for point-of-care glucose monitoring in professional healthcare settings. Collectively, these findings reinforce cobas® pulse as a clinically dependable and technologically advanced successor to the Inform II in professional healthcare environments.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Point-of-care Testing Device HemoCue® Glucose 201RT in Blood Glucose Estimation.Indian journal of community medicine : official publication of Indian Association of Preventive & Social Medicine · 2026
- Comparative Performance Analysis of Three Point-of-Care Glucose Testing Devices against a Central Laboratory Reference Method.EJIFCC · 2026
- B-030 Analytical Performance Evaluation of the Maccura Glycosylated Hemoglobin Analyzer G 02 (High-Performance Liquid Chromatography)Clinical Chemistry · 2026
- Glucometer Versus Standard Enzymatic Method for Gestational Diabetes Mellitus: A Systematic Review for Diagnostic Test Accuracy.International journal for quality in health care : journal of the International Society for Quality in Health Care · 2026
- A-360 Performance of rapid glucose and ketone screening methods in post-mortem vitreous fluid samples to determine ketoacidosis-mediated deathClinical Chemistry · 2026
- A Portable Electrical Response Verification System for Blood Glucose Meters Based on Dynamic Microcurrent Reproduction.Sensors (Basel, Switzerland) · 2026