Comparative Optical Performance of Two Enhanced Monofocal IOLs Across Pupil Sizes and Lens Powers.
Daniel Monsálvez-Romín, Susana Ortí-Navarro, César Antonio Albarrán-Diego, Nuria Garzón, María García-Montero
Department of Optics and Optometry and Vision Sciences, University of Valencia, Burjassot, Spain. · Department of Optometry and Vision, Complutense University of Madrid, Madrid, Spain.
Key points
From the paper's conclusion- Optical-bench testing indicates power- and pupil-dependent trade-offs between designs: the ISOPure is more robust to positive corneal spherical aberration at distance, whereas the RayOne EMV provides superior intermediate optical performance with greater pupil dependence.
Abstract
PURPOSE: To compare the optical performance of two enhanced monofocal intraocular lenses (IOLs) based on distinct strategies (isofocal power modulation vs controlled positive spherical aberration), across pupil diameters, corneal spherical aberration conditions, and lens powers. METHODS: Modulation transfer function (MTF) was measured in vitro with a PMTF bench (Lambda-X Ophthalmics) under International Organization for Standardization (ISO) 11979-2 conditions using two model corneas (aberration-free ISO1 and positive spherical aberration ISO2). Through-focus MTF (18.75 and 30 cycles per degree [cpd]), through-focus MTF area (0 to 18.75 and 0 to 30 cpd), and through-frequency MTF (0 to 60 cpd at 0 and -2.00 diopters [D]) were obtained for the ISOPure (BVI) and RayOne EMV (Rayner) at +10.00, +20.00, and +30.00 D, with 3- and 4.5-mm apertures. Aberration profiles were characterized with the NIMO TR1504 (Lambda X) by Zernike spherical aberration terms through the 8th order. RESULTS: Both IOLs showed a dominant distance-focus peak (0 D) across conditions. Increasing pupil size reduced peak MTF and MTF area for both designs. Under ISO1, the Ray-One EMV exhibited stable peak MTF across powers, whereas the ISOPure increased with power and slightly outperformed the RayOne EMV at +30.00 D. Under ISO2, the RayOne EMV showed a larger reduction in distance MTF while maintaining a broader through-focus profile and higher MTF at -2.00 D. NIMO measurements revealed strong negative spherical aberration with higher-order contributions for the ISOPure and a peripheral increase in positive spherical aberration for the RayOne EMV. CONCLUSIONS: Optical-bench testing indicates power- and pupil-dependent trade-offs between designs: the ISOPure is more robust to positive corneal spherical aberration at distance, whereas the RayOne EMV provides superior intermediate optical performance with greater pupil dependence.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Interferometric analysis of diffractive multifocal intraocular lenses.Journal of cataract and refractive surgery · 2026
- In Vitro Optical Bench Assessment of an Extended Depth-of-Focus Intraocular Lens Based on Segmented Refractive Design.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Visual Performance of a Novel Intraocular Lens With Higher-Order Aspheric Design Optimized for Intermediate Vision.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Pupil Size Associations and Refractive Error Tolerance in Enhanced Monofocal Intraocular Lenses With Different Central Optic Designs.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- From Earlier to Updated Quadrifocal Intraocular Lens Designs: An Objective Optical Metrology Analysis.Journal of cataract and refractive surgery · 2026
- Prediction of Monocular and Binocular Visual Acuity for the Pseudophakic Patients Using Simulated Modulation Transfer Function.Clinical ophthalmology (Auckland, N.Z.) · 2026