Pupil Size Associations and Refractive Error Tolerance in Enhanced Monofocal Intraocular Lenses With Different Central Optic Designs.
Koji Komatsu, Yoichiro Masuda, Kei Sano, Ryo Ohira, Shingo Mitooka, Masanobu Iida, Kosuke Ichihara, Tetsutaro Oki,
Department of Ophthalmology, Jikei University School of Medicine, Minato-ku, Tokyo, Japan.
Key points
From the paper's conclusion- Both IOLs preserved distance acuity and provided good intermediate vision. The NSP3 showed higher distance-spectacle independence and lower sensitivity to small residual refractive error, whereas the Impress tended to yield better intermediate vision with smaller pupils. These findings may help individualized intraocular lens selection.
Abstract
PURPOSE: To compare postoperative visual performance of two enhanced monofocal intraocular lenses with different central-optic designs (Nexroad NSP-3; Nidek [NSP3] and Vivinex Impress; Hoya Surgical Optics [Impress]) focusing on pupil size associations and refractive error tolerance. METHODS: This retrospective interventional case series was conducted at the Department of Ophthalmology, Machida Municipal Hospital (Tokyo, Japan). Consecutive cataract surgeries targeting emmetropia were included. Eligibility criteria were preoperative corneal astigmatism less than 1.00 diopter (D) and no prior ocular surgery or other ocular disease. At 1 month postoperatively, corrected distance visual acuity (CDVA), uncorrected distance, intermediate, and near visual acuity (UDVA [5 m]/UIVA [70 cm]/UNVA [30 cm]), contrast sensitivity, pupil diameter, and patient-reported outcomes (spectacle independence, dysphotopsia, satisfaction) were recorded. Generalized estimating equation models were used for between-group comparisons and regression analyses to account for inter-eye correlation. RESULTS: A total of 108 eyes were analyzed (NSP3: 55; Impress: 53). At 1 month postoperatively, CDVA, UDVA, UIVA, and UNVA did not differ between groups; contrast sensitivity was comparable. Distance-spectacle independence favored the NSP3 (53/55 [96.4%] vs 44/53 [83.0%]; P = .043). Smaller photopic pupils tended to improve UIVA for the Impress (r = 0.268; P = .053), whereas no correlation was seen for the NSP3 (r = 0.049; P = .721). UDVA correlated with postoperative spherical equivalent in the Impress (r = -0.504; P = .0001) but not the NSP3 (r = -0.114; P = .406). The generalized estimating equation interaction model showed a significant IOL-by-postoperative spherical equivalent interaction for UDVA (P = .019). CONCLUSIONS: Both IOLs preserved distance acuity and provided good intermediate vision. The NSP3 showed higher distance-spectacle independence and lower sensitivity to small residual refractive error, whereas the Impress tended to yield better intermediate vision with smaller pupils. These findings may help individualized intraocular lens selection.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Visual Performance of a Novel Intraocular Lens With Higher-Order Aspheric Design Optimized for Intermediate Vision.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Intra-individual comparison of two non-diffractive enhanced monofocal intraocular lenses.Journal of cataract and refractive surgery · 2026
- Visual, refractive and patient-reported outcomes after two trifocal intraocular lens designs: a 12-month comparative cohort study.BMC ophthalmology · 2026
- Comparison of 12-Month Visual Outcomes and Patient-Reported Measures between the BVI Isopure Enhanced Monofocal IOL and the AcrySof IQ Monofocal IOL.Korean journal of ophthalmology : KJO · 2026
- Comparative Optical Performance of Two Enhanced Monofocal IOLs Across Pupil Sizes and Lens Powers.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Comparison of visual outcomes and optical performance between Eyhance ICB00 and Lentis Comfort LS-313 MF15 intraocular lenses.International journal of ophthalmology · 2026