In Vitro Optical Bench Assessment of an Extended Depth-of-Focus Intraocular Lens Based on Segmented Refractive Design.
Biyue Tu, Yanxia Tong, Zhanpeng Zhu, Haifeng Jiang, Lihan Luo, Yong Wang
Department of Ophthalmology, Wuhan Aier Eye Hospital, Wuhan.
Key points
From the paper's conclusion- The low-addition IOL demonstrated superior intermediate optical performance compared with the monofocal and high-addition IOLs and fulfilled the ISO depth of focus criterion for EDOF IOLs. The high-addition IOL demonstrated superior optical performance at near focus.
Abstract
PURPOSE: To assess the in vitro optical performance of a low-addition segmented refractive intraocular lens (IOL) and compare it with extended depth-of-focus (EDOF) standards, monofocal IOLs, and high-addition bifocal IOLs manufactured from the same material. METHODS: Three identical material IOLs were assessed, namely a monofocal lens (HD), a low-addition multifocal lens (+2.00 diopters [D], SBL-2), and a high-addition multi-focal lens (+3.00 D, SBL-3). An optical bench measured the through-focus modulation transfer function (MTF) curves for 3- and 4.5-mm apertures. MTF values at five focal points (0 to -3.00 D) were evaluated. The area under the MTF curve was used to determine simulated visual acuity and depth of focus, compared to International Organization for Standardization (ISO) EDOF standards. Strehl ratio (SR) and United States Air Force (USAF) target imaging offered further quantitative and qualitative analyses. RESULTS: Through-focus MTF analysis confirmed distinct peaks corresponding to each IOL's design. The monofocal IOL provided the best distance (0 D) performance, whereas the low-addition IOL demonstrated the highest MTF values at intermediate foci (-1.50 to -2.50 D) and the high-addition IOL was superior at near (-3.00 D). The low-add IOL's depth of focus exceeded that of the monofocal IOL by 0.90 D at 0.20 logarithm of the minimum angle of resolution (logMAR), meeting the ISO EDOF standard. The monofocal IOL had the highest SR values. The USAF image indicated that the low-addition IOL provides stable image quality at intermediate focus, whereas the high-addition IOL excels at near focus. CONCLUSIONS: The low-addition IOL demonstrated superior intermediate optical performance compared with the monofocal and high-addition IOLs and fulfilled the ISO depth of focus criterion for EDOF IOLs. The high-addition IOL demonstrated superior optical performance at near focus.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Extended Depth-of-Focus Intraocular Lens Based on Aberration-Neutral Technique.Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) · 2026
- Comparative Optical Performance of Two Enhanced Monofocal IOLs Across Pupil Sizes and Lens Powers.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Interferometric analysis of diffractive multifocal intraocular lenses.Journal of cataract and refractive surgery · 2026
- Visual Performance of a Novel Intraocular Lens With Higher-Order Aspheric Design Optimized for Intermediate Vision.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- From Earlier to Updated Quadrifocal Intraocular Lens Designs: An Objective Optical Metrology Analysis.Journal of cataract and refractive surgery · 2026
- Tolerance of residual myopic refractive errors in eyes with enhanced monofocal intraocular lenses with higher-order aspheric optics.Japanese journal of ophthalmology · 2026