Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy Eyes
Veronica Noya-Padin, Cesar Sanchez-Sellero, Lua Souto-Blanco, Juan F. Otero-Mouteira, Eva Yebra-Pimentel, Maria J. Giraldez
Key points
From the paper's conclusion- Therefore, the study’s aims were to analyze the repeatability and reproducibility of two multi-diagnostic platforms (Visionix 130+ and Pentacam HR) in the measurement of corneal characteristics, and to assess inter-device agreement.
Abstract
Assessing the validity and measurement performance of optical instrumentation is essential to ensure its clinical utility. Therefore, the study’s aims were to analyze the repeatability and reproducibility of two multi-diagnostic platforms (Visionix 130+ and Pentacam HR) in the measurement of corneal characteristics, and to assess inter-device agreement. Anterior and posterior corneal curvature, eccentricity and central corneal thickness were measured using both devices in 31 healthy subjects. Each participant was subjected to six measurements per device, divided equally between two observers (three per observer). Both devices showed moderate-to-high reliability for anterior parameters and central corneal thickness (ICC > 79.89). For posterior parameters, Pentacam HR maintained moderate-to-high reliability (ICC: 78.28–99.34), whereas Visionix 130+ revealed decreased reliability (ICC: 5.16–77.77). Regarding eccentricity, high reliability was observed for Pentacam HR (ICC: 96.55), while Visionix 130+ exhibited low reliability (ICC: 68.81). No agreement was found between the results provided by the two devices for most of the parameters assessed. It is concluded that both devices demonstrate good repeatability and reproducibility when assessing anterior corneal surface and central corneal thickness, with the exception of eccentricity measured by Visionix 130+. For posterior surface measurements, Pentacam HR demonstrates moderate-to-high reliability, while Visionix 130+ exhibits low reliability for most parameters. Finally, the results of both devices are not interchangeable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of Anterior Segment Biometric Parameters Measured by Pentacam and Sirius in Healthy Myopic Eyes.Photodiagnosis and photodynamic therapy · 2026
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- Agreement of corneal thickness measured by Scheimpflug, SS-OCT, and specular microscopy in Chinese myopia.BMC ophthalmology · 2026
- Anterior Capsule Contrast and Capsulorhexis Geometry Across Visualization Systems in Cataract Surgery: An Exploratory Image-Based Comparison.Journal of clinical medicine · 2026
- Inter-device Agreement of Device-Designated Chord Mu and Chord Alpha Measurements in Patients Undergoing Preoperative Cataract Evaluation.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026