Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.
Rong Zhao, Zhiyang Zhang, Suyu Wang, Ziyi Chen, Guangguang Wu, Siyuan Ding, Qin Jiang, Jiajun Li,
Department of Ophthalmology, Eye Hospital of Nanjing Medical University, Nanjing 210029, China.
Key points
From the paper's conclusion- This study aimed to evaluate the agreement and clinical interchangeability of Pentacam AXL WAVE with IOLMaster 700 and Pentacam.
Abstract
Background/Objectives: Accurate preoperative biometry is essential for refractive cataract surgery. Pentacam AXL WAVE is an emerging multimodal platform integrating ocular biometry, anterior segment imaging, and corneal assessment, yet its interchangeability with established devices remains unclear. This study aimed to evaluate the agreement and clinical interchangeability of Pentacam AXL WAVE with IOLMaster 700 and Pentacam. Methods: This retrospective cross-sectional study compared Pentacam AXL WAVE with IOLMaster 700 and Pentacam across 175 eyes of 105 cataract patients. Interdevice agreement was evaluated using intraclass correlation coefficients, concordance correlation coefficients, and Bland-Altman analysis with multiple outlier-handling strategies; generalized estimating equations were applied to identify clinical factors associated with interdevice differences. Results: Pentacam AXL WAVE demonstrated good agreement with IOLMaster 700 and Pentacam for axial length and anterior corneal curvature parameters, but showed weaker agreement for white-to-white diameter, total keratometry, total corneal astigmatic vectors, and several extended anterior segment parameters. Fewer clinical factors explained interdevice differences versus IOLMaster 700, whereas ocular surface or corneal abnormalities, refractive status, and cataract characteristics were associated with greater differences versus Pentacam. Conclusions: These findings indicate that Pentacam AXL WAVE has parameter-specific interchangeability boundaries and can meet some routine biometric needs; however, for eyes requiring high refractive precision or presenting with complex ocular conditions, interpretation should be based on the agreement characteristics of individual parameters.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of anterior segment parameters in cataract patients with different axial lengths using IOL Master and Pentacam.BMC ophthalmology · 2026
- Concordance of Cataract Biometry Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Optical Biometry.Bioengineering (Basel, Switzerland) · 2026
- Systematic Differences in Corneal Curvature and Power Measurements Between the IOLMaster 700 and the Anterion and Mapping Strategies for Cross-Device Use.Diagnostics (Basel, Switzerland) · 2026
- Inter-device Agreement of Device-Designated Chord Mu and Chord Alpha Measurements in Patients Undergoing Preoperative Cataract Evaluation.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Comparison of spectral-domain optical coherence tomography biometry measurements with partial coherence interferometry optical biometryBMC Ophthalmology · 2026
- The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026