Inter-device Agreement of Device-Designated Chord Mu and Chord Alpha Measurements in Patients Undergoing Preoperative Cataract Evaluation.
Karanpreet S Multani, Anushka Raj, Liam D Redden, Jascha Wendelstein, Osamah Mian, Achim Langenbucher, Kamran M Riaz
Dean McGee Eye Institute, University of Oklahoma, Oklahoma City, Oklahoma. · Institut für Refraktive und Ophthalmo-Chirurgie, Zurich, Switzerland.
Key points
From the paper's conclusion- Clinically relevant inter-device differences in magnitude and orientation of chord mu and chord alpha were observed, with systematic centroid shifts and directional dispersion. These device-specific offsets indicate that such measurements are not interchangeable across platforms and should be interpreted in the context of the measuring device when applied to surgical planning.
Abstract
PURPOSE: To evaluate inter-device agreement of device-designated chord mu and chord alpha measurements, defined as device-designated linear displacements between the Purkinje reflex and the pupil center (chord mu) or limbal center (chord alpha), by four common devices: IOLMaster 700 (Carl Zeiss Meditec AG), Lenstar 900 (Haag-Streit), iTrace (Tracey Technologies), and Pentacam (Oculus Optikgeräte GmbH). METHODS: This was a retrospective cross-sectional study of patients undergoing preoperative cataract evaluation at a tertiary academic center. Eyes with measurements from all four devices between 2020 and 2023 were included. Exclusion criteria included prior ocular trauma, surgery, or pathology affecting imaging. Chord mu and alpha were analyzed using bivariate methods, including centroid direction and inter-device comparisons. Biometry and corneal curvature were also analyzed. RESULTS: A total of 110 eyes from 77 patients (mean age: 63 ± 11 years) were included. Across devices, mean chord mu values were approximately 0.35 ± 0.20 mm and mean chord alpha values were approximately 0.45 ± 0.23 mm, with substantial variability in both magnitude and orientation. Most pairwise inter-device comparisons demonstrated statistically significant differences for both chord mu and chord alpha. The IOLMaster 700 consistently demonstrated temporal centroid displacement and the widest directional dispersion, whereas other devices showed predominantly nasal centroid orientation. Left and right eyes demonstrated similar patterns but were analyzed separately. CONCLUSIONS: Clinically relevant inter-device differences in magnitude and orientation of chord mu and chord alpha were observed, with systematic centroid shifts and directional dispersion. These device-specific offsets indicate that such measurements are not interchangeable across platforms and should be interpreted in the context of the measuring device when applied to surgical planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of anterior segment parameters in cataract patients with different axial lengths using IOL Master and Pentacam.BMC ophthalmology · 2026
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- Assessment of Chord Mu and Chord Alpha in Relation to Post-ICL Implantation Visual Quality.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Agreement of corneal thickness measured by Scheimpflug, SS-OCT, and specular microscopy in Chinese myopia.BMC ophthalmology · 2026
- Quantitative Ciliary Body Imaging Using Image-Registered Swept-Source AS-OCT and Ultrasound Biomicroscopy.Translational vision science & technology · 2026
- Systematic Differences in Corneal Curvature and Power Measurements Between the IOLMaster 700 and the Anterion and Mapping Strategies for Cross-Device Use.Diagnostics (Basel, Switzerland) · 2026