The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.
Malak Habib, Brett P Dyer, Shu-Kay Ng, Alfred K Lam, Nicholas H Andrew
School of Medicine and Dentistry, Griffith University, Gold Coast, Queensland, Australia. · Griffith Health, Griffith University, Gold Coast, Queensland, Australia.
Key points
From the paper's conclusion- In eyes undergoing cataract surgery, the Eye-star demonstrated superior repeatability to the IOLMaster for keratometry but notably inferior repeatability for most en face variables.
Abstract
PURPOSE: To compare the repeatability of and describe the agreement between the IOLMaster 700 (Carl Zeiss Meditec AG) and Eyestar 900 (Haag-Streit) under identical test conditions. METHODS: This was a comparative cross-sectional study conducted at a single-surgeon ophthalmology clinic in Australia. Adults presenting for cataract surgery underwent two scans each on both the IOLMaster and Eyestar in random order. Bland and Altman's methods were used to assess repeatability and agreement. RESULTS: A total of 160 left eyes satisfied inclusion. For anterior keratometry, the 95% repeatability (test-retest) errors for the Thibos power components were lower with the Eyestar (M = 0.23, J0 = 0.20, J45 = 0.15 diopters [D]) than the IOLMaster (0.28, 0.21, and 0.15 D). For posterior keratometry, the errors were also lower with the Eyestar (M = 0.05, J0 = 0.02, J45 = 0.03 D) than the IOLMaster (0.09, 0.07, 0.07 D). The segmental axial variables had comparable repeatability. White-to-white distance was more repeatable with the IOLMaster (0.14 mm) than the Eyestar (0.26 mm). The Eyestar had notably inferior repeatability for the chord of angle alpha (ICX = 0.51, ICY = 0.57 mm) compared to the IOLMaster (0.14, 0.24 mm). For the chord of angle kappa, the repeatability of the Eyestar (PCX = 0.48, PCY = 0.49 mm) was also inferior to that of the IOLMaster (0.17, 0.15 mm). The Eyestar consistently measured a lower posterior corneal power; the mean difference in the posterior flat curvature (PK1) was -0.40 D (95% CI: -0.42 to 0.40). The biometers had close agreement for the segmental axial variables. CONCLUSIONS: In eyes undergoing cataract surgery, the Eye-star demonstrated superior repeatability to the IOLMaster for keratometry but notably inferior repeatability for most en face variables.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Repeatability of an SD-OCT biometer (MS-39 AL) and agreement with a validated biometer (IOLMaster 700).Journal of cataract and refractive surgery · 2026
- Comparison of spectral-domain optical coherence tomography biometry measurements with partial coherence interferometry optical biometryBMC Ophthalmology · 2026
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- Concordance of Cataract Biometry Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Optical Biometry.Bioengineering (Basel, Switzerland) · 2026
- Systematic Differences in Corneal Curvature and Power Measurements Between the IOLMaster 700 and the Anterion and Mapping Strategies for Cross-Device Use.Diagnostics (Basel, Switzerland) · 2026
- Comparison of anterior segment parameters in cataract patients with different axial lengths using IOL Master and Pentacam.BMC ophthalmology · 2026