Anterior Capsule Contrast and Capsulorhexis Geometry Across Visualization Systems in Cataract Surgery: An Exploratory Image-Based Comparison.
Hisaharu Suzuki
Zengyo Suzuki Eye Clinic, Fujisawa 251-0871, Kanagawa, Japan.
Key points
From the paper's conclusion- Conclusions: Anterior capsule contrast, as captured in recorded surgical video, was objectively higher with NGENUITY than with conventional microscopy, whereas a modest improvement in capsulorhexis circularity was observed only with version 1.5.
Abstract
Introduction: Objective, image-based comparisons of anterior capsule visualization across conventional microscopy and digital heads-up visualization systems remain limited. This exploratory retrospective study quantitatively compared capsule contrast and capsulorhexis geometry among conventional microscopy and two successive generations of a digital heads-up system, both of which apply yellow digital filtering. Methods: This retrospective, non-randomized observational study included 67 eyes of 60 patients undergoing phacoemulsification using a conventional microscope (21 eyes) or NGENUITY with yellow filtering [version 1.4 (N1.4; 25 eyes) or 1.5 (N1.5; 21 eyes)], with visualization modality determined by clinical availability at the time of surgery. Color contrast ratio (CCR), ovality index, and concentricity were quantified from masked video analysis using linear mixed-effects models (patient as random intercept) with Holm-Bonferroni-adjusted comparisons. Results: Mean CCR was significantly higher in both NGENUITY groups than the Microscope group (N1.4: 1.55 ± 0.55; N1.5: 1.70 ± 0.29; Microscope: 1.14 ± 0.09; both p ≤ 0.008), with no difference between versions. The ovality index was significantly lower in the N1.5 group than the Microscope group (2.90 ± 1.78% vs. 5.08 ± 3.38%; p = 0.022), though this difference did not reach significance in the per-patient sensitivity analysis (p = 0.071); no significant difference was observed between the Microscope and N1.4 groups. Concentricity did not differ significantly among groups. Conclusions: Anterior capsule contrast, as captured in recorded surgical video, was objectively higher with NGENUITY than with conventional microscopy, whereas a modest improvement in capsulorhexis circularity was observed only with version 1.5. These exploratory, hypothesis-generating findings are intended to motivate, rather than substitute for, a prospective, randomized, within-system study; they provide a standardized measurement framework showing that contrast parameters differ across visualization modalities, while geometric advantages across system generations warrant confirmation in such future studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy EyesPhotonics · 2026
- Comparison of Treatment Zone Decentration Between VISUMAX 500 and VISUMAX 800 Keratorefractive Lenticule Extraction Using Anterior Segment Optical Coherence Tomography.Journal of clinical medicine · 2026
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- Regional Corneal Densitometry Changes and Endothelial Stability After Cruciate-Pattern Nd:YAG Laser Posterior Capsulotomy: A Retrospective Self-Controlled Study.Diagnostics (Basel, Switzerland) · 2026
- Comparison of Cumulative Dissipated Energy Between Two Phacoemulsification Systems Using a Modified Flip-and-Chop Technique With Multidirectional Jerking Movement From a Phacoemulsification Handpiece.Cureus · 2026
- Comparison of early clinical and optical quality-related outcomes after SMILE using VisuMax 800 versus VisuMax 500 for myopia and low-to-moderate astigmatism in a Chinese population: a single-center retrospective studyFrontiers in Medicine · 2026