Comparison of Anterior Segment Biometric Parameters Measured by Pentacam and Sirius in Healthy Myopic Eyes.
Han Song, Wei Yang, Xiaoxue Wang, Jiani Wei, Jianwen Wang
Heilongjiang Provincial Eye Hospital, Harbin, Heilongjiang, China; Harbin Bright Eye Hospital, Harbin, Heilongjiang, China. · Harbin Bright Eye Hospital, Harbin, Heilongjiang, China.
Key points
From the paper's conclusion- Pentacam and Sirius measurements showed high correlation and good agreement and can provide complementary clinical references. However, systematic biases were observed in the absolute ACD, Kf, Ks, and AST values obtained with the two devices. Device-specific reference ranges are therefore advisable.
Abstract
PURPOSE: To compare the correlation and agreement between measurements of anterior segment biometric parameters obtained with Pentacam and Sirius in healthy myopic eyes. METHODS: A total of 679 patients with myopia aged 17-49 years (mean ± SD, 23.37 ± 5.93 years) were enrolled, including 431 males (63.5%) and 248 females (36.5%). The spherical equivalent (SE) of the right eye ranged from -0.63 D to -14.75 D (mean ± SD, -4.44 ± 1.84 D). Central corneal thickness (CCT), anterior chamber depth (ACD), flat keratometry (Kf), steep keratometry (Ks), and corneal astigmatism (AST) were measured with each device. Paired t-tests or Wilcoxon signed-rank tests, correlation analysis, intraclass correlation coefficient (ICC) analysis, and Bland-Altman analysis were used to assess correlation and agreement between the two systems. RESULTS: CCT did not differ significantly between the two devices (p > 0.05), whereas ACD, Kf, Ks, and AST differed significantly (all p < 0.05). Pentacam yielded lower ACD, Kf, and AST values but a higher Ks value than Sirius. All parameters were significantly and positively correlated (r = 0.929-0.984, all p < 0.001). All ICC values exceeded 0.92 (lower limits of the 95% CIs >0.91, all p < 0.001), indicating excellent agreement. Bland-Altman plots showed that the differences in all parameters clustered closely around zero, with narrow 95% limits of agreement. CONCLUSION: Pentacam and Sirius measurements showed high correlation and good agreement and can provide complementary clinical references. However, systematic biases were observed in the absolute ACD, Kf, Ks, and AST values obtained with the two devices. Device-specific reference ranges are therefore advisable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Repeatability, Reproducibility and Inter-Device Agreement of Anterior and Posterior Corneal Parameters Measured with Visionix 130+ and Pentacam HR in Healthy EyesPhotonics · 2026
- Concordance of Corneal Tomographic Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Scheimpflug Tomography.Bioengineering (Basel, Switzerland) · 2026
- Comparison of anterior segment parameters in cataract patients with different axial lengths using IOL Master and Pentacam.BMC ophthalmology · 2026
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- Agreement of corneal thickness measured by Scheimpflug, SS-OCT, and specular microscopy in Chinese myopia.BMC ophthalmology · 2026
- Quantitative Ciliary Body Imaging Using Image-Registered Swept-Source AS-OCT and Ultrasound Biomicroscopy.Translational vision science & technology · 2026