Evaluating Reliability and Validity of Heart Rate from the Garmin Instinct 2 During Deadlift Exercises at Different Intensities.
Jamaal K Bovell, Setareh Zarei, Kynlee Sperry, Jose Vasquez, Jeff Guzman, Catalina Kerwin, Juni Timmins, Nathaniel G Bodell,
Department of Kinesiology and Nutrition Science, University of Nevada, Las Vegas, Las Vegas, NV 89154, USA. · Department of Kinesiology and Outdoor Recreation, Southern Utah University, Cedar City, UT 84720, USA.
Key points
From the paper's conclusion- This study evaluated the accuracy and reliability of wrist-worn Garmin Instinct 2 monitors during deadlift exercise across multiple intensities, using the Polar H10 chest strap as the criterion measure.
Abstract
Background and Objectives: Wearable heart rate monitors are widely used for performance monitoring and safety, yet their validity during resistance training is limited. This study evaluated the accuracy and reliability of wrist-worn Garmin Instinct 2 monitors during deadlift exercise across multiple intensities, using the Polar H10 chest strap as the criterion measure. Methods: Twenty-one recreationally active adults (four women, 17 men, 0 identified otherwise) completed deadlift sets at low (RPE 3-4) and moderate-to-high (RPE 6-8) intensities while heart rate (HR) was recorded using four Garmin Instinct 2 devices placed on the inner and outer wrists of each forearm and one Polar H10 chest strap. Inter-device reliability was examined using the Intraclass Correlation Coefficient (ICC [1,2]) and the Coefficient of Variation (CV). Validity was assessed via mean absolute percentage error (MAPE) and Lin's Concordance Correlation Coefficient (CCC). Results: Outer wrist placement met the threshold for reliability and one measure of validity during low intensity exercise (average HR CV = 2.7%, ICC = 0.88, MAPE = 8.5% [left forearm], 9.5% [right forearm], CCC = 0.55 [left forearm], 0.52 [right forearm]; maximal HR CV = 2.9%, ICC = 0.88, MAPE = 3.2% [left forearm], 7.7% [right forearm], CCC = 0.61 [left forearm], 0.60 [right forearm]). No validity thresholds were met during moderate-to-high intensity exercise (average HR MAPE = 15.9% [left forearm], 15.9% [right forearm], CCC = 0.57 [left forearm], 0.60 [right forearm]; maximal HR MAPE = 30.1% [left forearm], 13.5% [right forearm], CCC = 0.53 [left forearm], 0.55 [right forearm]). Inner wrist placement did not meet any reliability or validity threshold for any condition. Conclusions: Outer wrist placement of the Garmin Instinct 2 achieved acceptable MAPE under low-intensity loading, but measures of accuracy appeared to decline with increased intensity. Inner wrist placement failed to return any conditions in which reliability or validity thresholds were met for either average or maximal heart rate. For training or occupational settings requiring precise HR monitoring, chest strap monitors or bicep-worn sensors are still recommended.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Criterion Validity of Smartwatches for Measuring Heart Rate and Energy Expenditure During Resistance Exercise.Bioengineering (Basel, Switzerland) · 2026
- Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test-Retest Experiment Among Healthy Young Men.Bioengineering (Basel, Switzerland) · 2026
- Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation StudyJMIR Formative Research · 2026
- Pulse vs. R-R Interval: Validity and Reliability of the Garmin Elevate v4 Sensor for Ultra-Short HR and HRV Monitoring in Healthy Adults.Sensors (Basel, Switzerland) · 2026
- Method comparison of a ring-based PPG device and a chest-strap monitor for heart rate and heart rate variability in shooting athletes.BMC sports science, medicine & rehabilitation · 2026
- Assessing the agreement and reliability of Magene® for heart rate monitoring.Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine · 2026