Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test-Retest Experiment Among Healthy Young Men.
Tae-Hyung Lee, Jong-Hee Bae, Jae-Myun Ko
Exercise Physiology Laboratory, Department of Physical Education, Yonsei University, Seoul 03722, Republic of Korea. · Department of Physical Education, Daegu National University of Education, Daegu 42411, Republic of Korea.
Key points
From the paper's conclusion- Therefore, early post-set error contained a reproducible session-level component, but individual 5 s readings were not interchangeable with ECG measurements.
Abstract
Smartwatch heart-rate (HR) displays may be inaccurate during rapid post-exercise transitions. This study evaluated timepoint-specific smartwatch-electrocardiography (ECG) error and its reproducibility across three resistance-exercise sessions. Sixty healthy young men completed three identical sessions five days apart using the Apple Watch Series 7, Galaxy Watch 4, Fitbit Charge 5, and Garmin Vivosmart 5. Displayed HR was recorded before each set and at 5, 30, and 60 s afterward. The primary outcome was participant-level session-mean error across eight 5 s observations. Mean absolute error increased in the range of 1.94-2.32 bpm before exercise to 8.71-25.43 bpm at 5 s and decreased to a range of 3.58-10.06 bpm at 30 s. Session-mean 5 s error intraclass correlation coefficients were 0.892, 0.771, 0.879, and 0.834 for Apple, Galaxy, Fitbit, and Garmin, respectively, indicating moderate-to-good through good-to-excellent reliability. Session effects were not statistically significant. Raw single-observation limits of agreement were substantially wider than session-mean limits, and all devices exhibited proportional bias. Participant-level variance estimates remained similar in exploratory models including ECG-response and wrist-position covariates. Therefore, early post-set error contained a reproducible session-level component, but individual 5 s readings were not interchangeable with ECG measurements. These findings support cautious interpretation of post-set HR displays without establishing a stable participant-specific error trait.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Criterion Validity of Smartwatches for Measuring Heart Rate and Energy Expenditure During Resistance Exercise.Bioengineering (Basel, Switzerland) · 2026
- Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation StudyJMIR Formative Research · 2026
- Evaluating Reliability and Validity of Heart Rate from the Garmin Instinct 2 During Deadlift Exercises at Different Intensities.Journal of functional morphology and kinesiology · 2026
- Evaluation of Blood Pressure and Heart Rate Measurements from Smartwatch HUAWEI WATCH D2 Data: A Preliminary Report.Healthcare (Basel, Switzerland) · 2026
- Performance evaluation of smartwatches: Can they match clinical standards for ECG analysis?Biomedical Signal Processing and Control · 2025
- Pulse vs. R-R Interval: Validity and Reliability of the Garmin Elevate v4 Sensor for Ultra-Short HR and HRV Monitoring in Healthy Adults.Sensors (Basel, Switzerland) · 2026