Criterion Validity of Smartwatches for Measuring Heart Rate and Energy Expenditure During Resistance Exercise.
Jung-Hyun Yun, Jae-Myun Ko, Tae-Hyung Lee
Department of Sports Science, Konkuk University, Chungju 27478, Republic of Korea. · Exercise Physiology Laboratory, Department of Physical Education, Yonsei University, Seoul 03722, Republic of Korea.
Key points
From the paper's conclusion- This study evaluated the criterion validity of four commercial smartwatches (Apple Watch Series 8, Samsung Galaxy Watch 5, Fitbit Charge 5, and Garmin Vivosmart 5) for measuring heart rate (HR) and energy expenditure (EE) during resistance exercise.
Abstract
Commercial smartwatches are increasingly used to monitor physiological responses during exercise; however, their validity for resistance exercise remains uncertain. This study evaluated the criterion validity of four commercial smartwatches (Apple Watch Series 8, Samsung Galaxy Watch 5, Fitbit Charge 5, and Garmin Vivosmart 5) for measuring heart rate (HR) and energy expenditure (EE) during resistance exercise. Seventy healthy men aged 20-39 years performed two sets each of the bench press, T-bar row, squat, and deadlift at individualized 10-repetition maximum loads. Smartwatch-derived HR was evaluated against electrocardiography, whereas smartwatch-derived workout-session EE was evaluated against indirect calorimetry. The Apple Watch showed the lowest measurement error for mean HR across the protocol (bias = -0.36 bpm; MAE = 1.81 bpm; ICC = 0.969) and for overall dynamic HR (bias = -0.72 bpm; MAE = 4.95 bpm; ICC = 0.904). HR agreement deteriorated markedly immediately after each set, particularly for the Samsung and Fitbit devices, which underestimated ECG-derived HR by 25.60 and 19.59 bpm, respectively. In contrast, EE estimates showed poor agreement with indirect calorimetry across all devices (ICC = 0.021-0.033). These findings demonstrate that the agreement of smartwatch-derived HR with electrocardiography measurements varies across time points during resistance exercise and indicate that smartwatch-derived workout-session EE should be interpreted cautiously when accurate metabolic assessment is required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test-Retest Experiment Among Healthy Young Men.Bioengineering (Basel, Switzerland) · 2026
- Evaluating Reliability and Validity of Heart Rate from the Garmin Instinct 2 During Deadlift Exercises at Different Intensities.Journal of functional morphology and kinesiology · 2026
- Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation StudyJMIR Formative Research · 2026
- Performance evaluation of smartwatches: Can they match clinical standards for ECG analysis?Biomedical Signal Processing and Control · 2025
- Evaluation of Blood Pressure and Heart Rate Measurements from Smartwatch HUAWEI WATCH D2 Data: A Preliminary Report.Healthcare (Basel, Switzerland) · 2026
- Comparison of smartwatch-derived energy expenditure estimation during outdoor activities.Frontiers in digital health · 2026