Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation Study
Jasper Gielen, Catharina Nina Van Oost, Glen Debard, Romy Sels, Nele A. J. De Witte, Toon Colman, Bert Bonroy, Jean-Marie Aerts
KU Leuven · Thomas More University
Key points
From the paper's conclusion- We hypothesized that physical activity irregularity, cognitive stress, and thermal climate conditions would affect measurement accuracy.
Abstract
Background: Commercial wearable devices allow for continuous heart rate (HR) monitoring in daily life. Their accuracy under ecologically valid conditions, however, remains insufficiently independently tested, especially during irregular activity, cognitive stress, and variable climates. Objective: This study evaluated the HR accuracy of 10 commercially available wearables under controlled variations in physical activity, cognitive stress, and temperature. We hypothesized that physical activity irregularity, cognitive stress, and thermal climate conditions would affect measurement accuracy. Methods: Forty-five healthy adults (21-68, mean 34, SD 12 y) completed a standardized protocol in climate-controlled chambers simulating neutral (23 °C), hot (36 °C), and cold (10 °C) conditions. Tasks included rest, cognitive stress (Montreal Imaging Stress Task), steady walking, and intermittent walking. Each of the 10 devices (Fitbit Charge 6, Fitbit Inspire 3, Garmin Vivosmart 5, Garmin Vivoactive 5, Apple Watch SE, Google Pixel Watch 2, Polar Ignite 3, Polar Pacer, Xiaomi Watch 2, and Oura Ring Gen 3) was compared against electrocardiogram-derived HR from a Zephyr BioHarness chest strap. Accuracy was assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), repeated-measures concordance correlation coefficient (CCC), and Bland-Altman analysis. Results: Significant variability across the devices was observed. Fitbit Charge 6 (MAE 4.5 bpm, MAPE 5.5%, CCC 0.93) and Google Pixel Watch 2 (MAE 4.9 bpm, MAPE 6.7%, CCC 0.87) showed strong agreement with the gold standard. In contrast, Fitbit Inspire 3, Polar Ignite 3, Polar Pacer, and Oura Ring displayed larger errors (MAE 9-14 bpm, MAPE 11%-16%) and lower CCC values (0.45-0.66). The climate conditions did not significantly affect the measurement accuracy of the test devices. The activity type, however, did have a significant effect: intermittent walking increased errors for multiple devices. Conclusions: Wearable HR measurement accuracy is device-specific and context-dependent. Moderate climates did not impair performance, but irregular movement reduced accuracy. Fitbit Charge 6 and Google Pixel Watch 2 demonstrated the highest reliability, supporting their use in health and sports monitoring. Careful device selection and context-aware interpretation remain critical for applied and clinical applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement of consumer wearable pulse oximeters across activity conditions and skin pigmentation.Frontiers in digital health · 2026
- PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone EffectsSensors · 2026
- Criterion Validity of Smartwatches for Measuring Heart Rate and Energy Expenditure During Resistance Exercise.Bioengineering (Basel, Switzerland) · 2026
- Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test-Retest Experiment Among Healthy Young Men.Bioengineering (Basel, Switzerland) · 2026
- Evaluation of Blood Pressure and Heart Rate Measurements from Smartwatch HUAWEI WATCH D2 Data: A Preliminary Report.Healthcare (Basel, Switzerland) · 2026
- Method comparison of a ring-based PPG device and a chest-strap monitor for heart rate and heart rate variability in shooting athletes.BMC sports science, medicine & rehabilitation · 2026