Mila Alves Matos Rodrigues, Thiago Danillo Silva, Jessika Teodoro Santos, Vinnycius Nunes de Oliveira, Nicolas Evangelista Marques, João Victor Rosa de Freitas, Marilia Santos Andrade, Rodrigo Luiz Vancini,
Faculdade de Educação Física, Universidade Federal de Goiás, Goiânia, Brazil. · Departamento de Fisiologia, Universidade Federal de São Paulo, Brazil.
Key points
From the paper's conclusion- The Magene® shows acceptable agreement with Polar® H7 during exercise but may lack precision for resting HR monitoring in clinical settings.
Abstract
To evaluate the accuracy and reliability of the Magene® heart rate (HR) monitor compared to the Polar® H7 (validated device) during rest, maximal graded exercise test, and active recovery. A within-group study was conducted with 68 healthy adults (21 women). Participants completed two laboratory sessions using both monitors in random order. HR values were collected at rest, during the Ellestad maximal treadmill stress testing protocol, and in active recovery. Relative reliability was assessed using the two-way random effects intraclass correlation coefficient for agreement (ICC(2,1)). Absolute reliability was assessed using typical error (TE, in bpm and %), coefficient of variation (CV, %), and minimal detectable change (MDC, in bpm and %). Agreement was examined using Wilcoxon signed-rank tests, Bland-Altman plots, and Pearson's r or Spearman's rho, as appropriate. No significant HR differences were found between devices (p > 0.05). Mean biases, calculated as Polar - Magene, were 1.421 bpm at rest, 0.659 bpm during exercise, 0.103 bpm at maximal effort, and 0.569 bpm during active recovery. ICC(2,1) values ranged from 0.656 (rest) to 0.881 (maximal effort). TE ranged from 3.784 bpm (2.0%) to 6.470 bpm (8.7%), and MDC from 10.5 bpm (5.6%) to 17.9 bpm (24.2%), depending on the condition. Correlations ranged from 0.682 to 0.881. The Magene® shows acceptable agreement with Polar® H7 during exercise but may lack precision for resting HR monitoring in clinical settings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pulse vs. R-R Interval: Validity and Reliability of the Garmin Elevate v4 Sensor for Ultra-Short HR and HRV Monitoring in Healthy Adults.Sensors (Basel, Switzerland) · 2026
- Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation StudyJMIR Formative Research · 2026
- Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test-Retest Experiment Among Healthy Young Men.Bioengineering (Basel, Switzerland) · 2026
- Evaluating Reliability and Validity of Heart Rate from the Garmin Instinct 2 During Deadlift Exercises at Different Intensities.Journal of functional morphology and kinesiology · 2026
- Method comparison of a ring-based PPG device and a chest-strap monitor for heart rate and heart rate variability in shooting athletes.BMC sports science, medicine & rehabilitation · 2026
- Criterion Validity of Smartwatches for Measuring Heart Rate and Energy Expenditure During Resistance Exercise.Bioengineering (Basel, Switzerland) · 2026