BMC sports science, medicine & rehabilitationOpen accessPhotoplethysmogram

Method comparison of a ring-based PPG device and a chest-strap monitor for heart rate and heart rate variability in shooting athletes.

Luna Tao, Hao Sun, Yugao Li, Siufung Sze, Xinzhen Pei, Cheng Chen, Yujun Cai

School of Physical Education, Shanghai University of Sport, Shanghai, China. · Shenzhen Ninenovo Technology Limited, Shenzhen, People's Republic of China.

Key points

From the paper's conclusion
  • These findings support the use of finger-based PPG for this specific measurement context, while broader application to daily training, movement-rich tasks, or competition settings requires further validation.

Abstract

INTRODUCTION: Ring-based photoplethysmography (PPG) devices are increasingly used in athlete monitoring, but evidence remains limited for precision-sport settings and finger-worn form factors. This study examined device-to-device agreement between a ring-based PPG device and a chest-strap monitor for heart rate (HR) and time-domain heart rate variability (HRV) metrics in shooting athletes. METHODS: Sixty-four shooting athletes completed a 10-min seated recording initiated within 5 min after a routine scored training session. RingConn Gen2 and Polar H10 signals were acquired simultaneously using synchronized timestamps, refined by cross-correlation, and analyzed in 90-s windows after excluding the first and last 30 s. HR, the standard deviation of normal-to-normal intervals (SDNN), and the root mean square of successive differences (RMSSD) were derived from inter-beat intervals. Primary agreement analyses were performed using participant-level paired device means, with window-level summaries retained as secondary descriptive analyses. RESULTS: Participant-level agreement was strongest for HR (; intraclass correlation coefficient [ICC]; mean bias bpm; 95% limits of agreement: to bpm; mean absolute error [MAE] bpm). Agreement was also strong for SDNN (; ICC; mean bias 1.791 ms; limits of agreement: to 6.690 ms; MAE ms), with a small proportional bias. RMSSD showed good but weaker agreement (; ICC; mean bias 3.237 ms; limits of agreement: to 15.065 ms; MAE ms). DISCUSSION: Under seated, low-motion post-training recovery conditions, the ring-based device showed close agreement with the chest-strap monitor, particularly for HR and SDNN. RMSSD agreement remained good but had wider limits of agreement, indicating that small within-athlete changes should be interpreted more cautiously. These findings support finger-based PPG as a practical option for standardized post-training recovery monitoring, while further work should evaluate performance during daily training, movement-rich tasks, and competition.

Read the paper

Discussion