Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery.
Se An Oh, Sung Yeop Kim, Jae Won Park, Ji Woon Yea, Jaehyeon Park, Yoon Young Jo
Department of Radiation Oncology, Yeungnam University Medical Center, Daegu 42415, Republic of Korea.
Key points
From the paper's conclusion- Thus, we propose the exploratory institutional observation of three targets, beyond which EPID-based correction may be considered to achieve optimal portal-dose prediction agreement in HyperArc-based stereotactic radiosurgery.
Abstract
Background/Objectives: Accurate verification of single-isocenter multiple-target (SIMT) stereotactic radiosurgery is challenging owing to the complexity of multi-lesion delivery and the sensitivity of stereotactic dose gradients. We aimed to evaluate the efficacy of electronic portal imaging device (EPID)-based multileaf collimator (MLC) transmission and dosimetric leaf gap (DLG) correction for improving portal-dose prediction agreement in SIMT stereotactic radiosurgery (SRS) and to propose an exploratory target count-based institutional action level. Methods: This retrospective analysis included 112 consecutive patients treated with HyperArc™-based SRS (1-13 targets). Treatment plans were calculated using the Acuros XB algorithm for 6 MV flattening filter-free (FFF) beams. Two sets of MLC parameters were compared for portal-dose image prediction (PDIP): (1) standard parameters measured using an ion chamber; (2) EPID-derived corrected parameters. Portal-dose accuracy was evaluated using gamma index analysis with a 95% pass rate threshold. Results: The uncorrected method showed a strong negative correlation between target number and gamma passing rates, with complex plans (≥9 targets) dropping as low as 70%. EPID-based correction substantially improved dose agreement, yielding consistent passing rates above 98%, regardless of target number. Although uncorrected parameters remained within tolerance for plans with one to two targets, accuracy declined markedly starting at three targets. Conclusions: Our findings indicate that the EPID-based correction of MLC transmission and DLG mitigated cumulative modeling discrepancies in complex SIMT SRS. Thus, we propose the exploratory institutional observation of three targets, beyond which EPID-based correction may be considered to achieve optimal portal-dose prediction agreement in HyperArc-based stereotactic radiosurgery.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- End to end dosimetric and geometric verification of single isocenter multitarget stereotactic radiosurgery using an anthropomorphic phantom: external audit validation.Radiological physics and technology · 2026
- Evaluation of ExacTrac for vertical setup verification in HyperArc radiosurgery: a phantom and dosimetric study.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Impact of daily machine performance check on prediction models for patient‐specific quality assurance of multiple‐target stereotactic radiosurgeryMedical Physics · 2026
- Sensitivity in error detection of an electronic portal imaging device for patient-specific QA in head and neck VMAT plans.Radiological physics and technology · 2026
- A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy.Biomedical physics & engineering express · 2026
- Assessing Gamma Index Sensitivity to Selected Linear Accelerator Mechanical Errors Using Various Dosimetric Systems.Journal of clinical medicine · 2026