Evaluation of ExacTrac for vertical setup verification in HyperArc radiosurgery: a phantom and dosimetric study.
Suet Yi Kennis Ng, Yick Wing Ho, Ka Hei Jason Lok, Chi Yuen Harry Cheng
HKU Li Ka Shing Faculty of Medicine, The University of Hong Kong, Hong Kong. Electronic address: kennisnsy@connect.hku.hk. · Union Oncology Centre, Union Hospital, Hong Kong.
Key points
From the paper's conclusion- Therefore, ExacTrac may serve as a complementary intrafraction verification option for HyperArc, particularly during non-coplanar delivery, while CBCT remains important for volumetric anatomical assessment.
Abstract
HyperArc stereotactic radiosurgery (SRS) for multiple brain metastases (MBM) uses cone beam computed tomography (CBCT) for pretreatment setup, and megavoltage (MV) portal imaging for intrafraction verification. At gantry 0°, MV portal imaging has limited sensitivity to anterior-posterior motion corresponding to treatment-couch vertical displacement as this motion is approximately parallel to the imaging beam axis. This study evaluated stereoscopic kV ExacTrac as a complementary verification method and quantified the dosimetric effect of simulated vertical setup offsets. A Sun Nuclear STEEV cranial phantom was shifted from ± 1 to ± 10 mm. Lateral and longitudinal shifts were evaluated at couch angles 0°, 45°, 90°, 270°, and 315°, while vertical shifts were evaluated at couch angles 0° and additionally at 270°. Shift measurements from ExacTrac, MV portal imaging, and CBCT were compared where geometrically available. In 19 patients with 53 lesions, clinical plans were recalculated with vertical isocenter offsets from ± 0.5 to ± 3.0 mm in 0.5 mm increments. Changes in Paddick conformity index (Paddick CI), homogeneity index (HI), gradient index (GI), and doses to organs at risk (OARs) were evaluated. PTV size and distance from isocenter were assessed using Spearman correlations and linear mixed-effects modelling. At couch 0°, mean ExacTrac-CBCT discrepancies were 0.017 ± 0.201 mm laterally, 0.080 ± 0.101 mm longitudinally, and 0.008 ± 0.130 mm vertically. MV portal imaging did not report the introduced treatment-couch vertical shifts. At the tested non-coplanar couch angle of 270°, ExacTrac identified the introduced vertical shifts while MV remained insensitive to this displacement at gantry 0°. Simulated 0.5 mm vertical offsets reduced Paddick CI by approximately 3%, while 3 mm offsets reduced conformity by more than 50% and increased HI. Smaller PTVs showed progressively greater conformity loss as shift magnitude increased (shift × log10 PTV interaction β = -0.071, 95% CI -0.078 to -0.064, p < 0.001), whereas the shift × distance interaction was not significant (p = 0.919). OAR dose changes were small at the cohort level, generally less than 0.1 Gy. Under controlled phantom conditions, stereoscopic kV imaging provided additional sensitivity to treatment-couch vertical displacement relative to MV portal imaging at gantry 0°. Dosimetric simulations showed a size-dependent reduction in target conformity with increasing vertical offset. Therefore, ExacTrac may serve as a complementary intrafraction verification option for HyperArc, particularly during non-coplanar delivery, while CBCT remains important for volumetric anatomical assessment. Patient-based validation is required to establish minimum clinically detectable shifts and clinical outcome benefit.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- End to end dosimetric and geometric verification of single isocenter multitarget stereotactic radiosurgery using an anthropomorphic phantom: external audit validation.Radiological physics and technology · 2026
- Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery.Diagnostics (Basel, Switzerland) · 2026
- Beyond CBCT: real-time motion management in spine and non-spine bone SBRT using stereoscopic X-rays and surface guidance.Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2026
- Assessing Gamma Index Sensitivity to Selected Linear Accelerator Mechanical Errors Using Various Dosimetric Systems.Journal of clinical medicine · 2026
- How robust are secondary dose algorithms for Patient-Specific quality assurance in modulated stereotactic Radiosurgery?Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026
- A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy.Biomedical physics & engineering express · 2026