Sensitivity in error detection of an electronic portal imaging device for patient-specific QA in head and neck VMAT plans.
Chanchakorn Thaiwattana, Chirasak Khamfongkhruea, Tinnakorn Donmoon, Nuntawat Udee, Thunyarat Chusin, Titipong Kaewlek, Sumalee Yabsantia
Master of Science Program in Medical Physics, Department of Radiological Technology, Faculty of Allied Health Sciences, Naresuan University, Phitsanulok, 65000, Thailand. · Medical Physics Program, Princess Srisavangavadhana Faculty of Medicine, Chulabhorn Royal Academy, 10120, Bangkok, Thailand.
Key points
From the paper's conclusion- These findings provide preliminary evidence that EPID-based PSQA with SPC-derived tolerance limits can detect moderate-to-large systematic MLC opening errors under the evaluated 2D gamma-analysis conditions.
Abstract
Reliable detection of delivery errors is essential for patient-specific quality assurance (PSQA) in volumetric modulated arc therapy (VMAT), particularly for head-and-neck treatments involving complex multileaf collimator (MLC) modulation. This study evaluated the error-detection performance of a newly implemented electronic portal imaging device (EPID)-based PSQA system for head-and-neck VMAT plans. Statistical process control (SPC) was used to establish institution-specific tolerance limits for the 2%/2 mm, 3%/2 mm, and 3%/3 mm gamma criteria, yielding limits of 88.45%, 97.34%, and 98.48%, respectively. A separate set of ten clinically accepted treatment plans was modified using MATLAB to introduce systematic MLC opening errors of 1, 3, and 5 mm per MLC bank. All 3 and 5 mm errors were detected, whereas 1 mm errors showed sensitivities of 60%, 70%, and 10% for the 2%/2 mm, 3%/2 mm, and 3%/3 mm criteria, respectively. Specificity was 100% within the evaluated set of ten clinically accepted original treatment plans. These findings provide preliminary evidence that EPID-based PSQA with SPC-derived tolerance limits can detect moderate-to-large systematic MLC opening errors under the evaluated 2D gamma-analysis conditions. Larger studies incorporating broader delivery-error models and clinically relevant dose-based evaluation are warranted.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Assessing Gamma Index Sensitivity to Selected Linear Accelerator Mechanical Errors Using Various Dosimetric Systems.Journal of clinical medicine · 2026
- Statistical process control and error curve modelling identify superior pre-treatment verification metrics for nasopharyngeal carcinoma and cervical cancer VMAT: a retrospective studyFrontiers in Oncology · 2026
- Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devicesJournal of Applied Clinical Medical Physics · 2026
- A deep learning approach for EPID-based patient-specific quality assurance incorporating daily machine-state variations.Frontiers in oncology · 2026
- It’s not your device, it’s how you use it – Sensitivity and specificity of PSQA devices in the SEAFARER head and neck studyRadiotherapy and Oncology · 2026
- ROC-based evaluation of tight gamma criteria for detecting MLC and jaw positioning errors in brain SRS using SRS MapCHECK.Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) · 2026