Sensitivity in error detection of an electronic portal imaging device for patient-specific QA in head and neck VMAT plans.

Chanchakorn Thaiwattana, Chirasak Khamfongkhruea, Tinnakorn Donmoon, Nuntawat Udee, Thunyarat Chusin, Titipong Kaewlek, Sumalee Yabsantia

Master of Science Program in Medical Physics, Department of Radiological Technology, Faculty of Allied Health Sciences, Naresuan University, Phitsanulok, 65000, Thailand. · Medical Physics Program, Princess Srisavangavadhana Faculty of Medicine, Chulabhorn Royal Academy, 10120, Bangkok, Thailand.

Key points

From the paper's conclusion
  • These findings provide preliminary evidence that EPID-based PSQA with SPC-derived tolerance limits can detect moderate-to-large systematic MLC opening errors under the evaluated 2D gamma-analysis conditions.

Abstract

Reliable detection of delivery errors is essential for patient-specific quality assurance (PSQA) in volumetric modulated arc therapy (VMAT), particularly for head-and-neck treatments involving complex multileaf collimator (MLC) modulation. This study evaluated the error-detection performance of a newly implemented electronic portal imaging device (EPID)-based PSQA system for head-and-neck VMAT plans. Statistical process control (SPC) was used to establish institution-specific tolerance limits for the 2%/2 mm, 3%/2 mm, and 3%/3 mm gamma criteria, yielding limits of 88.45%, 97.34%, and 98.48%, respectively. A separate set of ten clinically accepted treatment plans was modified using MATLAB to introduce systematic MLC opening errors of 1, 3, and 5 mm per MLC bank. All 3 and 5 mm errors were detected, whereas 1 mm errors showed sensitivities of 60%, 70%, and 10% for the 2%/2 mm, 3%/2 mm, and 3%/3 mm criteria, respectively. Specificity was 100% within the evaluated set of ten clinically accepted original treatment plans. These findings provide preliminary evidence that EPID-based PSQA with SPC-derived tolerance limits can detect moderate-to-large systematic MLC opening errors under the evaluated 2D gamma-analysis conditions. Larger studies incorporating broader delivery-error models and clinically relevant dose-based evaluation are warranted.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers