End to end dosimetric and geometric verification of single isocenter multitarget stereotactic radiosurgery using an anthropomorphic phantom: external audit validation.
S Balasubramanian, D Anbalagan, P Thanasekar, Rashi Agrawal
Department of Radiation oncology, Max super speciality hospital Vaishali, W3, Sector 1, Vaishali, Ghaziabad, 201010, India. bals20@gmail.com. · Department of Radiation oncology, Max super speciality hospital Vaishali, W3, Sector 1, Vaishali, Ghaziabad, 201010, India.
Key points
From the paper's conclusion- Our study aimed to perform a comprehensive end-to-end (E2E) dosimetric and geometric evaluation of the newly commissioned Varian Edge linear accelerator (LINAC) with the HyperArc™ VMAT system for multi-target stereotactic radiosurgery (SRS) to ensure clinical accuracy and safety in the treatment of multiple brain metastases.
Abstract
Our study aimed to perform a comprehensive end-to-end (E2E) dosimetric and geometric evaluation of the newly commissioned Varian Edge linear accelerator (LINAC) with the HyperArc™ VMAT system for multi-target stereotactic radiosurgery (SRS) to ensure clinical accuracy and safety in the treatment of multiple brain metastases. E2E testing was conducted using the MAX-EA anthropomorphic phantom, incorporating ion chamber and film inserts. The phantom was immobilized with a Q-fix Encompass system and scanned with a 0.625 mm CT protocol. Five spherical PTVs range between (0.3-9.2 cc) were planned using Eclipse™ v18.01 with 6 MV-FFF beams and optimized via the Photon Optimizer algorithm with Acuros XB (1.25 mm grid). A HyperArc™ plan prescribing 15 Gy to each target was generated. Ionization chamber, Gafchromic EBT-XD film, portal dosimetry (EPID), and Multi-Target Winston-Lutz (W-L) tests were used to evaluate dosimetric and geometric accuracy. The measured-to-calculated dose difference was - 3.3% for the HyperArc plan. Film analysis yielded average gamma pass rates > 94.9% (2%/2 mm), > 96.5% (3%/1 mm) and > 98.68% (3%/2 mm). Portal dosimetry achieved > 96.05% average gamma pass (2%/2 mm). The W-L test all targets demonstrated localization accuracy within 0.5 mm and MPC measurements confirmed sub-millimeter isocenter accuracy within 0.29 mm. E2E evaluation of the Edge LINAC with HyperArc VMAT demonstrated submillimeter geometric precision and robust dosimetric accuracy. Also, the Results met AAPM-RSS benchmarks. While rigid phantom validation inherently idealizes real patient-specific anatomical and setup variations. These high-quality baseline results support the clinical accuracy of the treatment workflow.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- RADT-46 Evaluation of brain stereotactic radiosurgery plan quality using a dual-model deep learning framework: a cross-platform analysisNeuro-Oncology · 2026
- Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery.Diagnostics (Basel, Switzerland) · 2026
- Evaluation of ExacTrac for vertical setup verification in HyperArc radiosurgery: a phantom and dosimetric study.Medical dosimetry : official journal of the American Association of Medical Dosimetrists · 2026
- Impact of daily machine performance check on prediction models for patient‐specific quality assurance of multiple‐target stereotactic radiosurgeryMedical Physics · 2026
- A patient-specific Winston-Lutz system for multi-target, single isocentre radiotherapy.Biomedical physics & engineering express · 2026
- Coplanar VMAT versus HyperArc: a comparative study in brain metastases stereotactic radiosurgery.La Radiologia medica · 2026