Quality indicators and between-scan variability of SS-OCT-derived angle alpha and angle kappa in cataract patients: a retrospective analysis of clinically repeated scans
Chen Li, Daiquan Chen, Ruoyi Yao, Weihua Yang
Key points
From the paper's conclusion- Purpose To investigate the impact of quality indicators (QI) from swept-source optical coherence tomography (SS-OCT) biometry on the between-scan variability of angle alpha and angle kappa in cataract patients.
Abstract
Purpose To investigate the impact of quality indicators (QI) from swept-source optical coherence tomography (SS-OCT) biometry on the between-scan variability of angle alpha and angle kappa in cataract patients. Methods This retrospective consecutive case series included 1,346 cataract patients (2,692 scan records) examined using the IOLMaster 700. Eyes were grouped based on laterality and QI grades. Absolute differences in angle alpha, angle kappa, and white-to-white (WTW) distance between the two clinically repeated scans were analyzed. Results Eyes with a QI of “Warning” exhibited significantly greater variability in angle alpha and angle kappa compared to those with a QI of “Successful” (all P < 0.05). In contrast, QI had no significant influence on the variability of axial length, corneal curvature, anterior chamber depth, lens thickness, or central corneal thickness. Conclusion Suboptimal QI was associated with greater between-scan variability in angle alpha and angle kappa measurements in SS-OCT biometry. Under high-quality measurement conditions, both parameters remained stable. In clinical practice, greater emphasis should be placed on QI results, and values for angle alpha and angle kappa should be interpreted with caution when measurement quality is poor, to optimize surgical planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparison of anterior segment parameters in cataract patients with different axial lengths using IOL Master and Pentacam.BMC ophthalmology · 2026
- Comparison of spectral-domain optical coherence tomography biometry measurements with partial coherence interferometry optical biometryBMC Ophthalmology · 2026
- The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- Concordance of Cataract Biometry Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Optical Biometry.Bioengineering (Basel, Switzerland) · 2026
- Agreement of Anterior Segment Biometrics Measured Using Image-Registered Swept-Source OCT and High-Frequency Ultrasound Biomicroscopy.Journal of glaucoma · 2026