Agreement of Anterior Segment Biometrics Measured Using Image-Registered Swept-Source OCT and High-Frequency Ultrasound Biomicroscopy.
Lucas L Humayun, Anna Heur, Michie Namba, Alexander Ko, Galo Apolo, Qifa Zhou, Benjamin Y Xu
Roski Eye Institute, Keck School of Medicine, University of Southern California, Los Angeles, CA, USA. · Albany Medical College, Albany NY, USA.
Key points
From the paper's conclusion- The Insight 100 demonstrated excellent intra-device repeatability and good-to-excellent inter-device agreement with ANTERION measurements for most angle parameters. However, angle-based measurements, particularly in the nasal sector, were not interchangeable. These findings highlight the importance of device-specific interpretation when evaluating angle configuration across imaging modalities.
Abstract
PURPOSE: To assess the inter-device agreement of ocular biometric parameters, including angle configuration, measured using swept-source anterior segment OCT (AS-OCT) and high-frequency ultrasound biomicroscopy (UBM). METHODS: Patients aged 18 years and older were consecutively enrolled and scanned with both the ANTERION AS-OCT and Insight 100 UBM systems under standardized lighting conditions to assess inter-device reproducibility. A subset of participants underwent repeat imaging to assess intra-device repeatability. Images acquired along the horizontal (temporal-nasal) meridian were analyzed. Pupillary diameter (PD), anterior chamber depth (ACD), anterior chamber width (ACW), lens vault (LV), angular opening distance (AOD500), and scleral spur angle (SSA500) were measured using automated, manufacturer-provided scleral spur detection software for the ANTERION AS-OCT while an expert grader manually assigned scleral spur localization for the Insight 100. Intraclass correlation coefficients (ICCs) were calculated to quantify agreement. RESULTS: 86 eyes of 62 participants were included in the inter-device analysis. Inter-device agreement was excellent for all biometric parameters (ICC: 0.75-0.99) except nasal AOD500 (ICC: 0.72). Greater AOD500 and nasal sector location were independently associated with greater inter-device disagreement in AOD500 measurement (P<0.001) on multivariable linear regression analysis. Intra-device repeatability (N=28) was excellent for all parameters measured using the ANTERION (ICC: 0.91-1.00) and Insight 100 (ICC: 0.85-1.00). CONCLUSIONS: The Insight 100 demonstrated excellent intra-device repeatability and good-to-excellent inter-device agreement with ANTERION measurements for most angle parameters. However, angle-based measurements, particularly in the nasal sector, were not interchangeable. These findings highlight the importance of device-specific interpretation when evaluating angle configuration across imaging modalities. PRECIS: The Insight 100, an arcuate-scanning ultrasound device, demonstrated excellent intra- and inter-device agreement with ANTERION OCT. However, variability in nasal anterior segment angle measurements suggests that these devices are not fully interchangeable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative Ciliary Body Imaging Using Image-Registered Swept-Source AS-OCT and Ultrasound Biomicroscopy.Translational vision science & technology · 2026
- Comparison of Biometric Measurements Between Anterior Segment Hyperparallel and Swept-Source Optical Coherence Tomography.Translational vision science & technology · 2026
- Comparison of anterior segment parameters in cataract patients with different axial lengths using IOL Master and Pentacam.BMC ophthalmology · 2026
- Evaluation of axial length measurement and other biometric parameters using three different generations of biometers.Frontiers in medicine · 2026
- Repeatability and Agreement between novel SS-OCT VG2 and CASIA2 in Anterior Segment Parameters.Photodiagnosis and photodynamic therapy · 2026
- Extending a Standard SD-OCT Device to Ocular Biometry with Additional Software: Agreement and Clinical Applicability Compared with Ultrasound and Optical Biometry.Journal of clinical medicine · 2026