Comparison of anterior segment parameters in cataract patients with different axial lengths using IOL Master and Pentacam.
Zhi-Yong Tian, Peng Zhao, Zhong-Yan Li, Zhi Chen, Jing-Shang Zhang, Ying-Yan Mao, Xiu-Hua Wan, Meng Li
Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, No.1, Dong Jiao Min Xiang, Dong Cheng District, Beijing, 100730, China. · Shandong Second Medical University, Weifang, Shandong, 261000, China.
Key points
From the paper's conclusion- Inter-device measurement biases between the IOL Master 700 and Pentacam HR are remarkably stable and are not exacerbated by extreme AL. These inherent biases stem from differing optical principles rather than patient-specific ocular geometries. Given that the LoA widths consistently exceed clinical tolerance thresholds, these devices are not clinically interchangeable for individual precise surgical planning.
Abstract
BACKGROUND: To evaluate the agreement and clinical interchangeability of anterior segment parameters measured by Swept-Source Optical Coherence Tomography (SS-OCT, IOL Master 700) and Scheimpflug imaging (Pentacam HR) across a continuous spectrum of axial lengths (AL). METHODS: This retrospective study consecutively included 294 cataract patients (294 eyes) divided into six AL subgroups. Central corneal thickness (CCT), anterior chamber depth (ACD), white-to-white (WTW) distance, and mean keratometry (Km) were measured under undilated conditions by a single technician. Relative reliability was evaluated using Intraclass Correlation Coefficients (ICC). Absolute agreement was analyzed using Bland-Altman limits of agreement (LoA) and compared against predefined clinical tolerance thresholds. One-way Analysis of Variance (ANOVA) was performed to determine whether inter-device measurement discrepancies (Δ) varied systematically across AL subgroups. RESULTS: Relative reliability was excellent for ACD, CCT, and Km (ICCs > 0.88), but moderate for WTW (ICC = 0.641). Bland-Altman analysis revealed consistent systematic differences. The 95% LoA widths for CCT (31.32 μm), WTW (0.826 mm), and Km (1.274 D) significantly exceeded a priori clinical thresholds. The LoA width for ACD (0.233 mm) marginally exceeded the strict threshold. Crucially, ANOVA demonstrated no statistically significant variations in measurement discrepancies across all six AL subgroups (all P > 0.05). CONCLUSIONS: Inter-device measurement biases between the IOL Master 700 and Pentacam HR are remarkably stable and are not exacerbated by extreme AL. These inherent biases stem from differing optical principles rather than patient-specific ocular geometries. Given that the LoA widths consistently exceed clinical tolerance thresholds, these devices are not clinically interchangeable for individual precise surgical planning.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- Agreement of Anterior Segment Biometrics Measured Using Image-Registered Swept-Source OCT and High-Frequency Ultrasound Biomicroscopy.Journal of glaucoma · 2026
- Agreement of corneal thickness measured by Scheimpflug, SS-OCT, and specular microscopy in Chinese myopia.BMC ophthalmology · 2026
- Evaluation of axial length measurement and other biometric parameters using three different generations of biometers.Frontiers in medicine · 2026
- Inter-device Agreement of Device-Designated Chord Mu and Chord Alpha Measurements in Patients Undergoing Preoperative Cataract Evaluation.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Quantitative Ciliary Body Imaging Using Image-Registered Swept-Source AS-OCT and Ultrasound Biomicroscopy.Translational vision science & technology · 2026