Comparison of spectral-domain optical coherence tomography biometry measurements with partial coherence interferometry optical biometry
Konuralp Yakar, Gökhan Özgür
Çankaya University · Samsun University
Key points
From the paper's conclusion- However, significant discrepancies in keratometric and astigmatic measurements limit their interchangeability.
Abstract
PURPOSE: This study aimed to evaluate the agreement and interchangeability between a spectral-domain optical coherence tomography (SD-OCT)-based biometer (REVO FC) and a partial coherence interferometry (PCI) device (AL Scan). METHODS: This prospective comparative cross-sectional study included 60 right eyes of 60 patients aged 40-82 years, diagnosed with senile cataracts. Measurements of axial length (AL), anterior chamber depth (ACD), keratometry (K1, K2), and astigmatism vector components (J0, J45) were conducted using both devices. Paired comparisons were performed using appropriate statistical tests. Agreement was evaluated using Bland-Altman analysis, while correlation and reliability were assessed using Pearson correlation coefficients and interclass correlation coefficients (ICC). RESULTS: No statistically significant differences were observed in AL (p = 0.507) or ACD (p = 0.218), with excellent correlations (AL: r = 0.998; ACD: r = 0.945). Bland-Altman analysis demonstrated excellent agreement for AL measurements, with a mean difference (bias) of - 0.01 mm and narrow 95% limits of agreement (LoA: -0.245 to 0.225 mm). For ACD, the mean difference was 0.02 mm, with a moderately wider LoA (- 0.244 to 0.288 mm), indicating an acceptable but less robust agreement compared to AL. The keratometric values were significantly higher with REVO FC (K1: +0.76 D; K2: +0.94 D; both p < 0.001), indicating a systematic bias. Despite strong correlations (K1: r = 0.980; K2: r = 0.924), the Bland-Altman analysis revealed wide limits of agreement. J0 exhibited no significant difference (p = 0.889), whereas J45 differed significantly (p = 0.028). Both vector components demonstrated moderate agreement, with relatively wide limits of agreement. CONCLUSION: REVO FC and AL Scan demonstrate excellent agreement for AL and ACD measurements and may be used interchangeably for these parameters. However, significant discrepancies in keratometric and astigmatic measurements limit their interchangeability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Extending a Standard SD-OCT Device to Ocular Biometry with Additional Software: Agreement and Clinical Applicability Compared with Ultrasound and Optical Biometry.Journal of clinical medicine · 2026
- The Repeatability and Agreement of Two SS-OCT Optical Biometers in Adult Eyes Undergoing Cataract Surgery.Journal of refractive surgery (Thorofare, N.J. : 1995) · 2026
- Clinical Interchangeability Boundaries of Pentacam AXL WAVE in Preoperative Biometry for Cataract Surgery: An Agreement Study with IOLMaster 700 and Pentacam.Journal of clinical medicine · 2026
- Evaluation of axial length measurement and other biometric parameters using three different generations of biometers.Frontiers in medicine · 2026
- Concordance of Cataract Biometry Measurements Obtained via Swept-Source Optical Coherence Tomography Versus Optical Biometry.Bioengineering (Basel, Switzerland) · 2026
- Quality indicators and between-scan variability of SS-OCT-derived angle alpha and angle kappa in cataract patients: a retrospective analysis of clinically repeated scansFrontiers in Medicine · 2026