Assessing Visual Field Defects in Glaucoma Using the Perceptive Scotoma Finder.

Huma Khurshid, Haeyoung Lopilly Park, Chan Kee Park, Hyung Bin Hwang, Yong Chan Kim

College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea. · College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea. lopilly@catholic.ac.kr.

Key points

From the paper's conclusion
  • The Perceptive Scotoma Finder demonstrated reproducible, sector-specific detection of established glaucomatous visual field defects and showed moderate agreement with conventional visual field and RNFL measures. These findings support its feasibility and potential as a rapid, low-cost adjunctive tool for glaucoma assessment; however, further evaluation in early glaucoma and screening populations is required.

Abstract

PURPOSE: To evaluate the diagnostic performance of the Perceptive Scotoma Finder, a dynamic touchscreen-based tool, for detecting glaucomatous visual field defects and examining its structure-function associations. METHODS: From the initial cohort of 143 eyes (86 participants), one eye per participant was included to avoid inter-eye correlation bias, comprising 58 glaucomatous eyes and 28 control eyes. All participants underwent monocular testing with a 30 Hz flickering dot stimulus mapped onto the Humphrey Visual Field 24-2 grid. Sectoral detection scores were compared with Humphrey Visual Field indices and retinal nerve fiber layer (RNFL) thickness. Group differences were analyzed with Mann-Whitney U tests, correlations with Spearman's rho, and diagnostic accuracy with receiver operating characteristic analysis. Test-retest reliability of the mean sector score was assessed in a subset of 26 eyes, including 18 glaucoma patients and 8 control subjects, using a two-way random-effects model for absolute agreement. RESULTS: Perceptive Scotoma Finder detection proportions were markedly higher in glaucoma than in controls across all Garway-Heath sectors (e.g., IN: 85.7% vs 1.9%; all p < 0.001). Detection scores showed correlations with mean deviation (ρ = 0.29 to 0.50) and pattern standard deviation (ρ = 0.36 to 0.69; all p < 0.001), and with RNFL thickness in all sectors except the temporal sector (p = 0.252). Sensitivity ranged from 0.68 to 0.91 and specificity from 0.83 to 0.90 across sectors. Test-retest reliability, based on 26 eyes, was 0.76 (95% confidence interval 0.72-0.89) by ICC, and Bland-Altman analysis showed a mean bias of 0.00 with 95% limits of agreement from -0.24 to 0.24. Overall discriminative performance produced an area under the curve of 0.79 (95% Confidence interval 0.76-0.83). CONCLUSIONS: The Perceptive Scotoma Finder demonstrated reproducible, sector-specific detection of established glaucomatous visual field defects and showed moderate agreement with conventional visual field and RNFL measures. These findings support its feasibility and potential as a rapid, low-cost adjunctive tool for glaucoma assessment; however, further evaluation in early glaucoma and screening populations is required.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers