Clinical ophthalmology (Auckland, N.Z.)

Evaluation of the Perimouse Online Suprathreshold Visual Field Test for Glaucoma Screening in a Chinese Population: A Comparison with the Humphrey Field Analyzer.

Xiaoyuan Shen, Zidong Chen, Kaixin Tan, Yanmei Fan, Pingping Liu, Haorui Wang, Jing Samantha Pan, Yangfan Yang,

State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-Sen University, Guangdong Provincial Key Laboratory of Ophthalmology and Visual Science, Guangdong Provincial Clinical Research Center for Ocular Diseases, Guangzhou, People's Republic of China. · Department of Psychology, Sun Yat-Sen University, Guangzhou, Guangdong, People's Republic of China.

Observational

Key points

From the paper's conclusion
  • Perimouse demonstrated reasonable overall diagnostic accuracy for detecting HFA-defined visual field defects. Diagnostic performance was higher for moderate and advanced glaucomatous defects than for early defects, with variable point-wise agreement with HFA. Perimouse may serve as a rapid, adjunctive tool for preliminary functional screening of glaucomatous visual field loss.

Abstract

PURPOSE: To evaluate the diagnostic performance and spatial agreement of an online suprathreshold perimeter (Perimouse) compared with the Humphrey Field Analyzer (HFA) for detecting glaucomatous visual field defects. PATIENTS AND METHODS: In this prospective comparative study, 174 participants (283 eyes with open-angle glaucoma or healthy controls) underwent Perimouse suprathreshold program and HFA. Abnormal points (missed stimuli in Perimouse; points with P<0.5% on HFA total deviation map) were compared. Diagnostic performance was assessed using receiver operating characteristic (ROC) curve analysis. Point-wise performance used neural network-derived area under the curve (AUC) and Cohen's κ. RESULTS: Perimouse abnormal-point counts were correlated with HFA abnormal-point counts (r = 0.88, P < 0.001) and negatively correlated with HFA mean deviation (r = -0.80, P < 0.001). The intraclass correlation coefficient for abnormal-point counts between Perimouse and HFA was 0.93, and Bland-Altman analysis showed a mean bias of 0.73 abnormal points, with 95% limits of agreement ranging from -10.70 to 12.17. A cutoff of five abnormal points achieved 85.9% sensitivity and 97.9% specificity (AUC = 0.97), with performance improving with disease severity. Point-wise κ values ranged from 0.32 to 0.78 (AUC, 0.67-0.96). The test duration for Perimouse was 2.64 ± 1.83 minutes. CONCLUSION: Perimouse demonstrated reasonable overall diagnostic accuracy for detecting HFA-defined visual field defects. Diagnostic performance was higher for moderate and advanced glaucomatous defects than for early defects, with variable point-wise agreement with HFA. Perimouse may serve as a rapid, adjunctive tool for preliminary functional screening of glaucomatous visual field loss.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Evaluation of the Perimouse Online Suprathreshold Visual Field Test for Glaucoma Screening in a Chinese Population: A Comparison with the Humphrey Field Analyzer — Science Skim