Performance of 24-2C Visual Field Grid in Detecting Central Glaucomatous Defects: A Comparative Study with Microperimetry.
Sami Bounetta, Aymeric Barbarroux, Amina Rezkallah, Thibaud Mathis, Victor Vermot-Desroches, Benoît Allignet, Laurent Kodjikian, Philippe Denis
Service d'Ophtalmologie, Hôpital Universitaire de la Croix-Rousse, Hospices Civils de Lyon, 69317 Lyon, France. · Service de Radiothérapie, Centre Léon Bérard, 28 rue Laennec, 69673 Lyon, France.
Key points
From the paper's conclusion- 24-2C VF grid has a potential complementary role in glaucoma assessment for detecting central VF defects in glaucoma patients. Its structure-function correlation supports its use as a promising tool in routine clinical practice.
Abstract
PRECIS: Our study shows that the 24-2C VF test provides central functional information strongly correlated with 10-2 and MAIA measurements, with a shorter testing time. It may serve as a useful complementary tool in routine glaucoma assessment. PURPOSE: To compare performance of 24-2C visual field (VF) grid and microperimetry in detecting central defects in glaucoma patients, and to correlate perimetric parameters with anatomical alterations. MATERIALS AND METHODS: This cross-sectional, single-center study included 214 eyes (118 glaucoma, 96 healthy). Participants underwent VF testing using 24-2C SITA-Faster, MAcular Integrity Assessment (MAIA) microperimetry, and 10-2 SITA-Standard grids. Macular imaging, including ganglion cell-inner plexiform layer complex (GC-IPL) analysis, was performed using Cirrus OCT. The primary outcome was a comparison of global indices between 24-2C grid and microperimetry. Secondary outcomes included comparisons with 10-2 grid, structure-function correlation, test duration, and 24-2C grid performance based on glaucoma severity. RESULTS: 24-2C Mean Deviation (MD) correlated significantly (P<0.001) with MAIA average threshold (R²=0.72) and CMS (R²=0.83). Significant correlations were also observed between 24-2C and 10-2 grids for MD (R²=0.77), PSD (R²=0.65), and CMS (R²=0.87). 24-2C grid required a shorter testing time (P<0.001) than MAIA and 10-2 grids. GC-IPL analysis revealed a significant correlation with 24-2C grid (P<0.001), particularly in the superior nasal sector (r=0.631). 24-2C grid demonstrated a stronger correlation with microperimetry for CMS, with r=0.674 in early-stage glaucoma and r=0.651 in advanced-stage glaucoma. CONCLUSION: 24-2C VF grid has a potential complementary role in glaucoma assessment for detecting central VF defects in glaucoma patients. Its structure-function correlation supports its use as a promising tool in routine clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Beyond the Index: Association of AI-CO Central Image Area with Central Visual Field Damage in GlaucomamedRxiv · 2026
- Comparison of 24-2 and 24-2C Visual Field Testing in Eyes with Ocular Hypertension.Medicina (Kaunas, Lithuania) · 2026
- Compatibility of 24-2C central extra test points with OCT and OCTA parameters in early-stage glaucoma.International journal of ophthalmology · 2026
- Structure-function relationship between central visual field sensitivities and papillomacular bundle thickness measured using a single optical coherence tomography vertical B-scan in patients with glaucoma.Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie · 2026
- Online circular contrast perimetry performed in glaucoma patients in Sub-Saharan Africa: a multi-centre comparative study in Nigeria.Eye (London, England) · 2026
- Evaluation of the Perimouse Online Suprathreshold Visual Field Test for Glaucoma Screening in a Chinese Population: A Comparison with the Humphrey Field Analyzer.Clinical ophthalmology (Auckland, N.Z.) · 2026