Agreement of consumer wearable pulse oximeters across activity conditions and skin pigmentation.
Kailey Miller, Hayley Whitney, Natalia Chavez, Dhruv R Seshadri
Department of Mechanical Engineering, Lehigh University, Bethlehem, PA, United States. · Department of Bioengineering, Lehigh University, Bethlehem, PA, United States.
Key points
From the paper's conclusion- Discussion These findings demonstrate that agreement between wearable and clinical pulse oximeters varies by device and activity state, while no statistically significant pigmentation-associated effects were identified after correction for multiple comparisons.
Abstract
INTRODUCTION: Wearable pulse oximetry devices are increasingly used in consumer-grade health products and patient monitoring; however, concerns remain regarding their accuracy across diverse populations and non-ideal conditions. METHODS: This study evaluated the agreement of four consumer wearable pulse oximeters with two clinical comparator monitors in a cohort of 49 adults during seated rest, treadmill walking, and seated recovery conditions. Skin tone was assessed using both the visual Fitzpatrick scale and the Individual Typology Angle (ITA), derived from objective skin color measurements. Device performance was evaluated using signed and absolute measurement error, generalized estimating equations accounting for repeated measurements, and repeated-measures Bland-Altman analysis. RESULTS: Agreement varied by device and activity condition. Garmin showed a significant negative signed bias relative to the GE Dash during cooldown (-0.72 percentage points, 95% CI -1.19 to -0.25; Holm-adjusted p = 0.028), while Wellue showed a significant negative signed bias during treadmill activity (-1.03 percentage points, 95% CI -1.52 to -0.53; Holm-adjusted p < 0.001). Wellue also demonstrated significant condition-dependent changes between treadmill and cooldown for both signed and absolute error. No device-specific association between skin pigmentation and signed or absolute error remained significant after correction for multiple comparisons using either continuous ITA or Fitzpatrick grouping. Measurement availability was substantially reduced for Garmin during treadmill activity (82.4%) and was nearly absent for the Apple Watch during treadmill activity (0.4%). DISCUSSION: These findings demonstrate that agreement between wearable and clinical pulse oximeters varies by device and activity state, while no statistically significant pigmentation-associated effects were identified after correction for multiple comparisons. Incorporating objective skin tone metrics, diverse participant populations, and activity-aware evaluation protocols will be essential for improving device reliability, supporting equitable digital health technologies, and reducing disparities in wearable-based physiological monitoring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation StudyJMIR Formative Research · 2026
- PPG-Based Heart Rate Accuracy in Hispanic Adults with Fitzpatrick III–V Skin Tones: An Evaluation of Body Composition and Skin-Tone EffectsSensors · 2026
- The Representation of Different Populations in Studies Assessing the Validity of Consumer Wearable Photoplethysmography-Based Measurements: Scoping Review.JMIR mHealth and uHealth · 2026
- Accuracy and Reliability of Pulse Oxygen Saturation Measured Using Wrist-Worn Oximeters.International journal of sports medicine · 2026
- Photoplethysmography in Diverse Skin Tones: Evaluating Bias in Smartwatch Health MonitoringCureus · 2025
- Pulse Oximeter Performance and Skin Pigment: Comparison of 34 Oximeters Using Current and Emerging Regulatory FrameworksAnesthesia & Analgesia · 2026