Evaluation of Blood Pressure and Heart Rate Measurements from Smartwatch HUAWEI WATCH D2 Data: A Preliminary Report.
Octavio A Lecona, Vania J Martínez-Garcés, Bruno Estañol, Antonio Barajas-Martinez, Paola V Olguín-Rodríguez, Ruben Fossion, Ana Leonor Rivera
Instituto de Ciencias Nucleares, Universidad Nacional Autónoma de México, Mexico City 04510, Mexico. · Departamento de Neurofisiología Clínica, Instituto Nacional de Ciencias Médicas y Nutrición Salvador Zubirán, Mexico City 14080, Mexico.
Key points
From the paper's conclusion- This study aimed to evaluate BP and HR measurements obtained with the HUAWEI WATCH D2 smartwatch by comparison with standard reference devices under controlled postural conditions.
Abstract
Background/Objectives: Wearable devices offer a scalable approach to repeated cardiovascular assessment through ambulatory spot measurements and have demonstrated adequate performance in measuring heart rate (HR). However, their reliability for blood pressure (BP) assessment remains uncertain. This study aimed to evaluate BP and HR measurements obtained with the HUAWEI WATCH D2 smartwatch by comparison with standard reference devices under controlled postural conditions. Methods: A total of 30 participants (15 women and 15 men; age range, 21-88 years) were enrolled. One participant could not undergo smartwatch measurement because of wrist-size limitations, and measurement acquisition failed in another participant. Consequently, paired measurements obtained in the seated and upright positions were analyzed for 28 participants. Smartwatch-derived systolic BP (SBP), diastolic BP (DBP), and HR were compared with a digital sphygmomanometer, an analog sphygmomanometer, and a digital pulse oximeter. Agreement was assessed using percentage error and Bland-Altman analysis, complemented by linear regression to evaluate proportional bias. Results: In the seated position, the smartwatch showed agreement with percentage errors below 4% for SBP and DBP, with concordance coefficients indicating good agreement. In the upright position, accuracy decreased, with percentage errors up to 14%, increased variability, wider limits of agreement, and the presence of systematic and proportional discrepancies. Measurement performance was influenced by device placement, and isolated failures highlighted sensitivity to anatomical and technical factors. Conclusions: Smartwatch-based BP monitoring is feasible under controlled conditions but remains dependent on posture and measurement conditions. These findings support the use of wearable devices for scheduled spot-check cardiovascular assessment under controlled conditions and underscore the need for standardized measurement protocols before their implementation in ambulatory or long-term monitoring applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of a wrist-worn dual-sensor applanation tonometry device for blood pressure measurement under rest and stress conditionsEuropean Heart Journal · 2026
- Wearable, cuffless, and portable devices for blood pressure monitoring (2015-2025): a scoping review of technologies, calibration, clinical validation, and regulatory frameworks.Frontiers in digital health · 2026
- Assessing Heart-Rate Display Reliability and Agreement Error Between Four Smartwatches and an Electrocardiography Criterion After Resistance Exercise: A Three-Session Test-Retest Experiment Among Healthy Young Men.Bioengineering (Basel, Switzerland) · 2026
- Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation StudyJMIR Formative Research · 2026
- Opportunistically Detecting Signs of Hypertension on a Consumer Smartwatch.Nature · 2026
- Toward patient-centered blood pressure monitoring: validation of a novel low-pressure and rapid oscillometric method according to the ISO 81060-2:2018/Amd1:2020/Amd2:2024 standard.Hypertension research : official journal of the Japanese Society of Hypertension · 2026