Automated perimetry in pediatric neuro-ophthalmic conditions: virtual reality-based vs. standard tabletop platforms.

Jeffrey Liu, Nikita Piryani, Zara Saleem, Rizul Naithani, Sharon F Freedman, Bo Wang, Mays A El-Dairi

Department of Ophthalmology, Duke University Medical Center, 2351 Erwin Road, Durham, NC 27710. · Department of Ophthalmology, Duke University Medical Center, 2351 Erwin Road, Durham, NC 27710; Wilmer Eye Institute, Johns Hopkins Hospital, 600 North Wolfe Street, Baltimore, MD 21287.

Key points

From the paper's conclusion
  • VRF (HVF-equivalent algorithm) testing showed moderate to good concordance with tabletop HVF testing in pediatric neuro-ophthalmic patients, with shorter testing duration and high patient preference, supporting its potential clinical value.

Abstract

OBJECTIVE: The game-based virtual reality field (VRF) testing algorithm has shown promise in pediatric neuro-ophthalmic patients, but the faster Humphrey visual field (HVF) equivalent algorithm has not been studied in this population. The purpose of this study was to compare testing outcomes between tabletop HVF and VRF (HVF-equivalent algorithm) in a pediatric neuro-ophthalmic population. DESIGN: Prospective comparison of diagnostic perimetry testing modalities. PARTICIPANTS: Children (<18 years of age) with a known or suspected neuro-ophthalmic condition. METHODS: Consecutive pediatric patients presenting to one neuro-ophthalmic clinic between January 2025 and April 2026 had clinically indicated visual field testing first with tabletop HVF testing (24-2, Size III, SITA Fast) and then VRF in the HVF-equivalent algorithm (24-2, Size III, AVA Fast). Global parameters, field defect agreement via masked grading, and point-by-point agreement (including total deviation [TD] and pattern deviation [PD] plots) between modalities were evaluated. HVFs with false-negative/false-positive rates >35% or fixation losses >50% were considered unreliable, applying less stringent reliability criteria for this pediatric population. MAIN OUTCOME MEASURES: Agreement rate between HVF and VRF in both defect detection by region (quadrants, and center) and individual plot points (<0.5% probability defect threshold). Comparisons across patient characteristics, perimetry results, and testing modality (HVF vs. VRF). RESULTS: One hundred fourteen patients (210 eyes; mean age 11.4 ± 3.6 years) were enrolled. Ninety-two children (168 eyes) completed both HVF and VRF testing, including 128 reliable and 40 unreliable HVF examinations. Regional defect agreement between HVF and VRF was 65.9% on average, while point-by-point agreement rates for TD and PD plots were 76.6% and 83.7%, respectively. Moderate positive associations were observed between HVF and VRF for mean deviation (R2M = .398, P < .001) and pattern standard deviation (R2M = .576, P < .001). VRF was significantly faster than HVF (2.1 ± 0.17 minutes vs 4.4 ± 1.2 minutes; P < .001) and most queried patients (51/53; 96.2%) preferred VRF to HVF testing. CONCLUSION: VRF (HVF-equivalent algorithm) testing showed moderate to good concordance with tabletop HVF testing in pediatric neuro-ophthalmic patients, with shorter testing duration and high patient preference, supporting its potential clinical value.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers