Comparison of the Melbourne rapid fields-web device-independent online perimeter and the octopus 600 perimeter in glaucoma patients.

George Y X Kong, Algis Vingrys, Manisha Ambawat, Md Mahboob, Shakira Bano, Syril Dorairaj, Shibal Bhartiya

The University of Melbourne, Parkville, VIC, Australia. · Marengo Asia Hospitals, Gurugram, Haryana, India.

Observational · n=113

Key points

From the paper's conclusion
  • Introduction The purpose of this study is to evaluate the agreement, repeatability, and testing efficiency of Melbourne Rapid Fields-Web (MRF-Web), a browser-based, device-independent perimeter, compared with the Octopus 600 perimeter in glaucoma patients.

Abstract

INTRODUCTION: The purpose of this study is to evaluate the agreement, repeatability, and testing efficiency of Melbourne Rapid Fields-Web (MRF-Web), a browser-based, device-independent perimeter, compared with the Octopus 600 perimeter in glaucoma patients. METHODS: This retrospective, cross-sectional, single-center observational study included 113 patients undergoing routine visual field assessment for glaucoma or glaucoma suspicion at Marengo Asia Hospitals, Gurgaon, India. MRF-Web testing was performed on a calibrated 14-inch laptop with Hindi voice guidance. Agreement between MRF-Web and Octopus 600 for Mean Deviation (MD) and pattern-based global indices was assessed using intraclass correlation coefficients (ICCs) and Bland-Altman analysis. Test-retest repeatability was evaluated in 51 patients who underwent repeat MRF-Web testing. RESULTS: MRF-Web showed strong group-level concordance with Octopus 600 for MD, with an ICC of 0.88 (95% CI, 0.81-0.92), Bland-Altman bias of -0.78 dB, and 95% limits of agreement from -4.72 to 3.17 dB. Concordance for the pattern-based indices was moderate, with an ICC of 0.72 (95% CI, 0.58-0.81), bias of -0.59 dB, and 95% limits of agreement from -4.41 to 3.24 dB. MRF-Web demonstrated good test-retest concordance, with an ICC of 0.86 (95% CI, 0.69-0.94). Mean per-eye test duration was significantly shorter with MRF-Web than with Octopus 600 (232.9 ± 49.8 versus 342.4 ± 94.4 seconds). CONCLUSIONS: MRF-Web demonstrated strong group-level concordance with Octopus 600 MD measurements and shorter per-eye testing times in this predominantly mild-disease cohort. The limits of agreement indicate that measurements from the two platforms should not be regarded as directly interchangeable, and further study is required in moderate and advanced glaucoma.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers