Ten-year insert wear and the migration of uncemented tibial and femoral components with either conventional polyethylene or sequentially annealed highly crosslinked polyethylene: a blinded randomized controlled trial using radiostereometric analysis.
Ruud de Ridder, Bart L Kaptein, Bart G Pijls, Rob G H H Nelissen, Jantsje H Pasma, Rianne Oomen, Herman H Kaptijn
Reinier Haga Orthopaedic Centre, Zoetermeer, Netherlands. · Department of Orthopaedics, Leiden University Medical Centre, Leiden, Netherlands.
Key points
From the paper's conclusion- There was no discernible increased wear for the HXLPE group after five years, whereas the wear for the PE group continued. After the initial migration, the uncemented TKA components remain stable at the ten-year follow-up. The use of HXLPE inserts are safe, when used in uncemented TKA, when considering wear and component migration.
Abstract
AIMS: The aims of this randomized controlled trial (RCT) was to compare the long-term insert wear and migration of the tibial and femoral components between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). METHODS: A double-blinded RCT was conducted including 96 TKAs. The wear of the insert, and migration of both the tibial and femoral components, was measured with model-based radiostereometric analysis up to ten years postoperatively. Patient-reported outcome measures (PROMs) were collected preoperatively and at follow-up. RESULTS: There was no progression of (medial) wear (from 0.15 mm at five years to 0.13 mm at ten years) for the HXLPE group after five years, whereas the wear rate increased for the PE group (from 0.22 mm at five years to 0.46 mm at ten years). There was no clinically relevant difference in terms of migration of the tibial and femoral component or PROMs between the HXLPE and PE groups at ten years. Migration of the tibial component (maximal total point motion; MTPM) after the ten-year follow-up for the HXLPE group was 0.99 mm (95% CI 0.83 to 1.18), and for the PE group, was 1.05 mm (95% CI 0.87 to 1.24). Migration of the femoral component (MTPM) after the ten-year follow-up for the HXLPE group was 1.16 mm (95% CI 0.99 to 1.35), and for the PE group, MTPM was 1.04 mm (95% CI 0.87 to 1.22). CONCLUSION: There was no discernible increased wear for the HXLPE group after five years, whereas the wear for the PE group continued. After the initial migration, the uncemented TKA components remain stable at the ten-year follow-up. The use of HXLPE inserts are safe, when used in uncemented TKA, when considering wear and component migration.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral stems: secondary analysis of a randomized controlled trial using radiostereometric analysis (RSA) with 10 years' follow-up.Acta orthopaedica · 2026
- Improved Wear Performance of Highly Cross-Linked Versus Conventional Polyethylene in Dual Mobility Cups.Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2026
- Association of polyethylene type with revision and dislocation risk following primary total hip arthroplasty: a population-based cohort study from the Danish Hip Arthroplasty Register.Acta orthopaedica · 2026
- The 10- Year Clinical and Radiographic Outcomes of a Vitamin E-Infused Highly Cross-Linked Polyethylene Cup in Total Hip Arthroplasty: A Multicenter Cohort Study.The Journal of arthroplasty · 2026
- Volumetric Wear of Highly Cross-Linked Polyethylene Sockets in Total Hip Arthroplasty Differs Among Brands: A Retrieval Study.Journal of functional biomaterials · 2026
- Wear Scar Morphology and Distribution in Retrieved Unicondylar Knee Arthroplasty Implants: Influence of Patient and Implant Factors.Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2026