Improved Wear Performance of Highly Cross-Linked Versus Conventional Polyethylene in Dual Mobility Cups.

Cristina Cocor, Agathe Sellier, Cyril Imbert, Sonia Ramos-Pascual, Julien Wegrzyn

Department of Orthopedic Surgery and Trauma, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland. · Dedienne Santé, Mauguio, France.

Key points

From the paper's conclusion
  • In conclusion, vit E XLPE DM mobile components had significantly reduced wear compared to UHMWPE DM mobile components.

Abstract

The purpose of the present study was to compare wear between vitamin E infused highly cross-linked polyethylene (vit E XLPE) versus ultra-high molecular weight polyethylene (UHMWPE) dual-mobility (DM) mobile components for total hip arthroplasty (THA), in worst-case conditions. Four in vitro tests were performed: (1) global wear test with pristine mobile components, (2) global wear test with accelerated ageing mobile components, (3) outer bearing only wear test with pristine mobile components, and (4) outer bearing only wear test with pristine mobile components and third-body particles. For each of these four tests, three vit E XLPE and three UHMWPE DM mobile components were tested, with the largest available DM cup size. Wear was calculated using the gravimetric method, adjusting for fluid uptake. Total wear at 5 million cycles was significantly lower for vit E XLPE compared to UHMWPE for all four tests: 79.3 ± 9.6 mg vs. 138.2 ± 7.2 mg (p = 0.003) for global wear with pristine mobile components, 54.9 ± 3.0 mg vs. 592.1 ± 79.6 mg (p = 0.007) for global wear with accelerated ageing mobile components, 59.8 ± 10.2 mg vs. 231.6 ± 15.0 mg (p < 0.001) for outer bearing wear with pristine mobile components, and 75.4 ± 17.5 mg vs. 370.5 ± 46.1 mg (p = 0.009) for outer bearing wear with pristine mobile components and third-body particles. In conclusion, vit E XLPE DM mobile components had significantly reduced wear compared to UHMWPE DM mobile components. Difference in wear was greatest for accelerated ageing and third-body particles tests, which may be most relevant for young and active patients, as well as for revision THA cases. Clinical testing of DM systems is necessary to confirm these findings in vivo.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Improved Wear Performance of Highly Cross-Linked Versus Conventional Polyethylene in Dual Mobility Cups — Science Skim