A randomized study on migration of the Spectron EF and the Charnley flanged 40 cemented femoral stems: secondary analysis of a randomized controlled trial using radiostereometric analysis (RSA) with 10 years' follow-up.
Håkon Greve Johannessen, Geir Hallan, Ove Furnes, Anne Marie Fenstad, Stein Håkon Lygre, Kristin Haugan, Thomas Kadar, Mona Badawy,
Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen, Norway. haakjohannessen@gmail.com. · Department of Clinical Medicine, Faculty of Medicine, University of Bergen, Bergen; The Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen; Coastal Hospital in Hagavik, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway.
Key points
From the paper's conclusion- After 10 years, the Spectron EF stem showed less retroversion, but more subsidence than the Charnley stem, with otherwise similar migration and clinical outcomes. XLPE articulating stems had more subsidence than CPE with the Spectron EF stem.
Abstract
BACKGROUND AND PURPOSE: Our primary aim was to compare the 10-year migration of the proximally roughened Spectron EF stem with the Charnley flanged 40 stems in cemented THA. Secondary aims were to assess the influence of polyethylene type on Spectron EF migration, and clinical outcomes. METHODS: In this randomized trial (NCT00698672), 150 patients (70% women; mean age 70 years) underwent cemented THA with either a Charnley stem (n = 30) or a Spectron EF stem (n = 120, 4 groups with different articulating surfaces). The articulations utilized conventional polyethylene (CPE) or highly crosslinked polyethylene (XLPE) cup liners combined with either cobalt-chromium (CoCr) or Oxinium (OxZr) femoral heads. RSA was performed postoperatively and up to 10 years. Migration was analyzed using linear mixed-effects models. Clinical outcome was assessed with the Harris Hip Score (HHS). RESULTS: After 10 years, 48 patients (14 Charnley, 34 Spectron EF) had evaluable RSA data. Both stems showed posterior translation, subsidence, and retroversion. Retroversion was greater for the Charnley stem than the Spectron EF stem at 3.2° (95% confidence interval [CI] 2.7-3.8) and 1.5° (CI 1.2-1.8), respectively, with a mean difference of 1.7° (CI 1.1- 2.4). The Spectron EF stem had more subsidence at 0.72 mm (CI 0.56-0.88) vs 0.32 mm (CI 0.06-0.57) for the Charnley stem, with a difference of 0.41 mm (CI 0.11-0.71). Spectron EF stems articulating with XLPE had 0.44 mm (CI 0.12-0.76) more subsidence than stems articulating with CPE. HHS improved from 41 (SD 15) to 86 (SD 19) for the Charnley stem and 45 (SD 13) to 86 (SD 15) for the Spectron EF stem with no statistically significant difference between the groups. CONCLUSION: After 10 years, the Spectron EF stem showed less retroversion, but more subsidence than the Charnley stem, with otherwise similar migration and clinical outcomes. XLPE articulating stems had more subsidence than CPE with the Spectron EF stem.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ten-year insert wear and the migration of uncemented tibial and femoral components with either conventional polyethylene or sequentially annealed highly crosslinked polyethylene: a blinded randomized controlled trial using radiostereometric analysis.The bone & joint journal · 2026
- Similar 2 year migration patterns for a shortened and a distally reduced stem design in uncemented total hip arthroplasty: a randomized controlled trial using radiostereometric analysis.Acta orthopaedica · 2026
- Ten-year migration analysis of a cementless metaphyseal-anchored short femoral stem in total hip arthroplasty.European journal of orthopaedic surgery & traumatology : orthopedie traumatologie · 2026
- The Triple-Tapered Uncemented Femoral Stems in Slight Varus Demonstrated Excellent Early Survivorship and Clinical Results.The Journal of arthroplasty · 2026
- Do modular tapered long stems differ in stability and function? A prospective comparison of two stems in complex femoral revision THAJournal of Orthopaedics and Traumatology · 2026
- Cementless calcar-guided short stems: 24-month clinical outcomes and migration analysis.Scientific reports · 2026