Abstract
Outcomes after total hip arthroplasty (THA) remain limited by wear and wear particle reaction. Although highly cross-linked polyethylene (HXLPE) mitigated this issue, there are major differences among the various brands available. This study quantified, for the first time, the volumetric wear on the retrievals from various HXLPE brands, using an optical coordinate measuring machine. Patient- and implant-related factors and component orientation were considered. As of 30.06.2025, 104 HXLPE sockets were available, and 79 could be analysed, excluding those too severely deformed/damaged. The median (range) time in situ was 8.7 (2.1-22.1) years. Wear volume and volumetric wear rate were 91 (1-674) mm3 and 9 (0-130) mm3/year, respectively. Highcross (n = 56, Medacta) experienced approximately five times more wear (p < 0.001) than Durasul (n = 17, Zimmer Biomet). The other brands (n = 6) were too small in group size at this stage for statistical analysis. The correlation (R2 = 0.94) between linear and volumetric wear was excellent, but high wearing retrievals had a strong impact on this analysis, whereas low wear retrievals showed high variability. Differences in wear were mostly driven by the brand, while other parameters had only a minor impact. This is the first retrieval study comparing volumetric wear among various brands of retrieved HXLPE bearings. Not all HXLPE bearings should be considered equal. As the cohort grows, further parameters and brands may be integrated.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Improved Wear Performance of Highly Cross-Linked Versus Conventional Polyethylene in Dual Mobility Cups.Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2026
- Association of polyethylene type with revision and dislocation risk following primary total hip arthroplasty: a population-based cohort study from the Danish Hip Arthroplasty Register.Acta orthopaedica · 2026
- Ten-year insert wear and the migration of uncemented tibial and femoral components with either conventional polyethylene or sequentially annealed highly crosslinked polyethylene: a blinded randomized controlled trial using radiostereometric analysis.The bone & joint journal · 2026
- Long-term results of a vitamin E-blended polyethylene liner in total hip arthroplasty: a minimum ten year computer aided design based evaluation of wear and component positioning.International orthopaedics · 2026
- Comparative outcomes of delta-on-delta and delta-on-highly cross-linked polyethylene bearings in total hip arthroplasty: a nationwide cohort study.Archives of orthopaedic and trauma surgery · 2026
- The 10- Year Clinical and Radiographic Outcomes of a Vitamin E-Infused Highly Cross-Linked Polyethylene Cup in Total Hip Arthroplasty: A Multicenter Cohort Study.The Journal of arthroplasty · 2026