Wear Scar Morphology and Distribution in Retrieved Unicondylar Knee Arthroplasty Implants: Influence of Patient and Implant Factors.
Carolina S Haboba Nebel, Ajay S Potluri, Aditya S Yadav, Anne Debenedetti, Spencer Schulte, Jennifer L Wright, Deborah J Hall, E Bailey Terhune,
Rush University Medical Center, Chicago, Illinois, USA.
Key points
From the paper's conclusion- This study quantitatively characterized the morphology and spatial distribution of wear scars in 95 surgically retrieved low-conformity UKA tibial inserts to identify patient and implant factors that drive wear magnitude and location.
Abstract
Unicondylar knee arthroplasty (UKA) utilization is expanding rapidly, and polyethylene wear remains a primary driver of long-term failure in low-conformity designs. This study quantitatively characterized the morphology and spatial distribution of wear scars in 95 surgically retrieved low-conformity UKA tibial inserts to identify patient and implant factors that drive wear magnitude and location. The tibial liners were evenly distributed between males and females, 88.5% were medial, median time in situ was 7 years (range: 2-21.5 years), median age at implantation was 61.5 years (range: 35.1-77.5 years), and ten designs were represented. High-resolution digital microscopy images analyzed through ImageJ were utilized to quantify wear scar area and wear scar location. Median wear scar area occupied around one-third of the articular surface. Multivariable Generalized Linear Models confirmed that larger wear areas were independently predicted by younger age at implantation (p ≤ 0.02) and longer time in situ (p ≤ 0.04). Among high-volume designs (n > 15) in the medial side-only cohort, historical conventional polyethylene showed significantly larger wear areas than highly cross-linked polyethylene (53% vs. 30%; p = 0.004), and larger centroid shifts (15.9% vs. 10.6%; p = 0.01), indicating more decentralized loading. This retrieval study shows that polyethylene degradation is therefore multifactorial: younger age at implantation accelerates wear magnitude, while time in situ, articular design and material type dictate wear propagation and tracking stability. Younger patients undergoing medial UKA represent a high-risk cohort; selecting modern implant designs with highly cross-linked polyethylene and geometric profiles that minimize peripheral centroid shifts may reduce edge loading and improve long-term survivorship.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ten-year insert wear and the migration of uncemented tibial and femoral components with either conventional polyethylene or sequentially annealed highly crosslinked polyethylene: a blinded randomized controlled trial using radiostereometric analysis.The bone & joint journal · 2026
- Coronal alignment-related mediolateral load distribution and polyethylene wear in a medial-pivot total knee arthroplasty design.Archives of orthopaedic and trauma surgery · 2026
- Associations between hospital characteristics, volume, and reasons for revision: a cohort study of 48,029 unicompartmental knee arthroplasties with 3,397 revisions from the Dutch Arthroplasty RegisterActa Orthopaedica · 2026
- Preoperative varus ≥10° does not compromise outcome after medial unicompartmental knee arthroplasty, whereas postoperative joint-line convergence angle and insert thickness influence functional results: A matched-pair analysis.Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA · 2026
- Fixed-bearing unicompartmental vs. total knee arthroplasty in elderly patients with medial osteoarthritis and anterior cruciate ligament deficiency: a retrospective comparative study.Frontiers in surgery · 2026
- Navigation-Assisted and Robotic Total Hip, Knee, and Unicompartmental Arthroplasty: Reconciling Positioning Accuracy with Biomaterial Performance and Long-Term Implant Survival.Bioengineering (Basel, Switzerland) · 2026